C Category Options and Descriptions
C01 Total Method Defines the method for calculating the cumulative flow. Options include:
Positive Flow
Negative Flow
Net Flow
Changing this method will reset the cumulative flow data. Default setting: Positive Flow.
POS
0
Default
NET
1
NEG
2
After selecting a command, confirm whether to Save or Not Save.
C02 Del. Total (Delete Total) Clears the cumulative flow data. Selecting this option will reset all cumulative flow values to zero.
No
0
Yes
1
After selecting a command, confirm whether to Save or Not Save.
C03 Dosing Mode Manually adjusts the dosing mode to either Count Up (incremental) or Count Down (decremental). Default setting: Off.
OFF
0
Count Up
1
Default
Count Down
2
After selecting a command, confirm whether to Save or Not Save.
C04 Dosing Target Sets the target value for dosing operations. Default value: 1000.
Data Type
float32
Default
1000
Minimum
0.1
Maximum
999,999,999
After selecting a command, confirm whether to Save or Not Save.
C05 Dosing Timer Resets the dosing timer. Minimum value: 1 second. Default value: 300 seconds.
Data Type
int
Default
300
Minimum
1
Maximum
3600
After selecting a command, confirm whether to Save or Not Save.