FU-ES EchoSense Instruction Manual
  • ⚠Safety Instructions
  • This instruction manual version is v1.0
    • Manual Update Log
  • Measurement principle and usage scenario
    • Firmware Update Date and Version Feature Description
  • Warranty
  • Exterior, Specs, Model
  • Installation and Setup
    • Mechanism Installation
    • Mechanical and Electrical iInstallation
    • Parameter setup
      • Z02 Automatic Pipe Size Detection Guide
  • Human Machine Interface
    • Introduction of human machine interface
    • Hybrid Page Setup and Usage Interface Guide
    • Monitoring Screen
    • Simulation Mode
    • Data interpretation on monitoring screens
    • Password input page
    • Setup screen
  • Communication
    • OCT output
    • Analog current output (0-20mA, 4-20mA)
    • RS485 and Modbus RTU Setting
    • Modbus RTU Protocol
      • Read-only memory read instruction(04) and memory lists
      • Read-Write Memory List
      • Read/Write Memory Read Command (03)
      • Read-write memory single Register write instruction(06)
      • Read/Write Memory Dual-Register Write Command (30)
      • Read/Write Memory Four-Register Write Command (31)
      • Single Bit Data Read/Write Commands (01/05) (Commonly Used in HMIs)
  • Parameters Table
    • Z Category Options and Descriptions
    • A Category Options and Descriptions
      • A02 Preset Pipe Specification Table for Different Models
      • A03 ~ A07 Data Range Explanation Based on Model Specifications
    • B Category Options and Descriptions
    • K Category Options and Descriptions
    • C Category Options and Descriptions
    • D Category Options and Descriptions
    • E Category Options and Descriptions
    • F Category Options and Descriptions
    • H Category Options and Descriptions
    • Pagasfsae 1
    • Page 1
  • Alarm and description of system conditions
  • Troubleshootings and Abnormalities
    • Iout 4-20/0-20mA debug process
    • OCT Alarm Debug Process
    • OCT Pulse Debug Process
    • RS485 Debug Process
  • Certification
  • FU-ES Product Page
  • LORRIC English Website
  • LORRIC Integration and Application Center
Powered by GitBook
On this page
  • Instantaneous flow volume and deviation
  • Flow speed
  • Cumulative flow speed
Export as PDF
  1. Human Machine Interface

Data interpretation on monitoring screens

PreviousSimulation ModeNextPassword input page

Last updated 1 year ago

Instantaneous flow volume and deviation

The maximum possible deviation for LORRIC’s flow meter is 0.5 % F.S. + 2.5 % O.R. which could be converted to the relation of maximum possible deviation and displayed data under different flow speed as seen below.

Flow speed

The data displayed on the screen is the flow speed in the T-connector. In most cases it will be a bit faster than the flow speed of the same ordinary pipe at the same flow volume.

Cumulative flow speed

The accumulated flow rate continuously counts until manually reset.

The maximum limit for the accumulated flow rate is 999,999,999 Liters, and it will reset to zero once exceeded.