OCT output
Specs
Please see the page of the monitoring screen.
Cable wiring
Please see the page of mechanical and electrical installation.
Parameter setup
Please refer to the parameter set-up page.
Set-up steps as below:
Please set up the basic parameters first.
Connect cable first according to the cable wiring instructions on this page.
Set up parameter E01 for selecting NC(Normal Close) or NO(Normal Open) when there is no signal.
Set up parameter E02 for specifying output signal content.
For selecting “pulse output”
Use E03 to confirm each pulse width.
Use F03 to simulate specific pulse output frequency at OCT for testing with external devices.
For selecting “pulse v2”
Use E03 to confirm each pulse width.
Use E14 to determine each pulse capacity.
F03 to simulate specific pulse output frequency at OCT for testing with external devices.
For selecting “out of range flow volume notice”.
Use D05 to turn on the alarm.
Use D06 D07 for specifying normal flow range.
Use D08 for specifying hysteresis range percentage. D07 is a positive value, it means the single range must exceed in order to disable the alarm.
Set up F01 and F02 to simulate flow volume and go back to the monitoring screen to confirm if the alarm works.
Turn off the simulation mode.
For selecting “Total Alarm.” (“Inform of over total target”)
Use D09 to turn off alarm.
Use D10 for specifying notification conditions. "Larger Than" is for positive flow direction, "Less Than" is for negative flow direction.
Use D11 for specifying total target, positive number is for positive flow, negative number is for negative flow.
Set up F01 and F02 to simulate flow volume and go back to the monitoring screen to confirm the alarm works.
Turn off the simulation mode.
Go back to the monitoring screen, finish.
Pulse output
The user uses parameter E13 to select a volume first, and the device outputs a digital square wave every time it accumulates through a specified volume.