FU-ES EchoSense Instruction Manual
  • ⚠Safety Instructions
  • This instruction manual version is v1.0
    • Manual Update Log
  • Measurement principle and usage scenario
    • Firmware Update Date and Version Feature Description
  • Warranty
  • Exterior, Specs, Model
  • Installation and Setup
    • Mechanism Installation
    • Mechanical and Electrical iInstallation
    • Parameter setup
      • Z02 Automatic Pipe Size Detection Guide
  • Human Machine Interface
    • Introduction of human machine interface
    • Hybrid Page Setup and Usage Interface Guide
    • Monitoring Screen
    • Simulation Mode
    • Data interpretation on monitoring screens
    • Password input page
    • Setup screen
  • Communication
    • OCT output
    • Analog current output (0-20mA, 4-20mA)
    • RS485 and Modbus RTU Setting
    • Modbus RTU Protocol
      • Read-only memory read instruction(04) and memory lists
      • Read-Write Memory List
      • Read/Write Memory Read Command (03)
      • Read-write memory single Register write instruction(06)
      • Read/Write Memory Dual-Register Write Command (30)
      • Read/Write Memory Four-Register Write Command (31)
      • Single Bit Data Read/Write Commands (01/05) (Commonly Used in HMIs)
  • Parameters Table
    • Z Category Options and Descriptions
    • A Category Options and Descriptions
      • A02 Preset Pipe Specification Table for Different Models
      • A03 ~ A07 Data Range Explanation Based on Model Specifications
    • B Category Options and Descriptions
    • K Category Options and Descriptions
    • C Category Options and Descriptions
    • D Category Options and Descriptions
    • E Category Options and Descriptions
    • F Category Options and Descriptions
    • H Category Options and Descriptions
    • Pagasfsae 1
    • Page 1
  • Alarm and description of system conditions
  • Troubleshootings and Abnormalities
    • Iout 4-20/0-20mA debug process
    • OCT Alarm Debug Process
    • OCT Pulse Debug Process
    • RS485 Debug Process
  • Certification
  • FU-ES Product Page
  • LORRIC English Website
  • LORRIC Integration and Application Center
Powered by GitBook
On this page
  • Alarm - A Major Abnormality
  • Alarm - B Out of Specification
  • Alarm - C Maintenance Required
  • Alarm - D Function Check
  • System Status
Export as PDF

Alarm and description of system conditions

Alarm - A Major Abnormality

Suggestions of system abnormality are the same with Failed message level mentioned in NAMUR NE107 standard.

The Output Display when the flow sensing function is abnormal or the host is abnormal.

“Error” is displayed on the LED screen, and the LCD backlight blinking red light.

Alarm - A Following explains error message bits from higher bit (7th digit) to lower bit (0th digit):

Bit

Definition Description of Errors

7

na

6

na

5

na

4

The flowmeter is not on the pipeline (off pipe)

3

Iout (0/4-20mA) IC Failure=1. Parameter D14 must be in the “on” or “do not disturb” state. (FW2.2.0 new function)

2

Iout (0/4-20mA) IC Overheating =1. Parameter D14 must be in the “on” or “do not disturb” state. (FW2.2.0 new function)

1

Abnormal pipe diameter sensing element.

0

Abnormal communication of ultrasonic components.

Alarm - B Out of Specification

Indicates out-of-range usage of flowmeter. Suggestions of system abnormality are the same with Out of Specification message level mentioned in NAMUR NE107 standard.

Indicates out of designed range flow volume and other matters needing attention. Could be set up in the main class of parameter D.

Display on the LED screen remains the same, and the LCD backlight will have a continuous red light.

Alarm - B Following explains error message bits from higher bit (7th digit) to lower bit (0th digit):

Bit

Definition Description of Errors

7

na

6

Cumulative volume exceeds target.

5

Data is larger than the corresponding upper limit of the analog output current.

4

Data is smaller than the corresponding lower limit of the analog output current.

3

Flow volume is higher than 6m/s.

2

(When OCT is set to Pulse output) Pulse width is too big, causing overlapping.

1

Instantaneous flow volume is higher than the defined upper limit.

0

Instantaneous flow volume is less than the defined lower limit.

Alarm - C Maintenance Required

Indicates maintenance requirement of flowmeter. Suggestions of system abnormality are the same with Maintenance Required message level mentioned in NAMUR NE107 standard.

Display on the LED screen remains the same, and the LCD backlight will have a continuous red light.

Alarm - C Following explains error message bits from higher bit (7th digit) to lower bit (0th digit):

Bit

Definition Description of Errors

7

na

6

na

5

na

4

na

3

Poor ultrasonic signal (0 bars)

2

The flowmeter installation problem leads to poor signal, it is recommended to lock the flowmeter evenly to achieve the best signal quality.

1

Iout(0/4-20mA) Excessive line impedance or no wiring = 1, normal = 0. Parameter D14 must be in the “on” or “do not disturb” state.

0

na

Alarm - D Function Check

Suggestions of system abnormality are the same with the Function Check message level mentioned in NAMUR NE107 standard.

The bit will be 1 when the device LCD screen is not one of the monitoring screens.

When the LED screen displays the set-up menu, the LCD backlight will have orange lighting.

Alarm - D Following explains error message bits from higher bit (7th digit) to lower bit (0th digit):

Bit

Definition Description of Errors

7

na

6

na

5

na

4

na

3

na

2

na

1

na

0

The device is not in the monitoring screen.

(It will become 1 in simulation mode and Setup screen)

System Status

Following explains status bits from higher bit (7th digit) to lower bit (0th digit):

Bit

Definition Description of Errors

7

na

6

na

5

na

4

na

3

na

2

na

1

na

0

OCT is conducting (Close)=1, OCT is not conducting (Open)=0

Pulse output status is not included.