Parameter setup

  1. After the mechanism is installed, execute Z03 to automatically initialize the flowmeter parameters, and the machine can work normally.

Basic tuning

  1. If there is still large flow data with the fluid stationary, execute a Z04 Zero Offset correction with water in the pipe and the fluid stationary.

  2. If there is a reference value for the flow rate, you can use the Z05 Scale Factor to multiply the flow rate data by a certain ratio to correct it to the correct value.

For advanced tuning settings, please refer to the complete parameter description page