A Category Options and Descriptions
Last updated
Last updated
Automatically detects pipe specifications and, upon completion, automatically switches to A02 to display the results. This function is identical to Z02.
Select a pipe model to adjust the data, or choose "Custom" to define pipe dimensions and material. Supported pipe specifications are listed below: A02 Different Model Preset Pipe Specification Table.
Custom
0
Default
cns w
1
cns e
2
jisk
3
astm sch 40 pvc
4
astm sch 40 sus
5
astm sch 80 pvc
6
jisg
7
iso pp
8
iso pvdf
9
After selecting the command, please confirm whether to Save or Not.
Specifies the unit for pipe dimensions, synchronized with B06. A03 to A07 follow the model data range.
mm
0
Default
inch
1
After selecting the command, please confirm whether to Save or Not.
Data Type
float32
Default
32
Minimum
1 mm
Maximum
200
After selecting the command, please confirm whether to Save or Not.
Data Type
float32
Default
Minimum
3.15 mm
Maximum
800
After selecting the command, please confirm whether to Save or Not.
Data Type
float32
Default
2.5
Minimum
0.1 mm
Maximum
50
After selecting the command, please confirm whether to Save or Not.
Data Type
float32
Default
Minimum
1 mm
Maximum
199
After selecting the command, please confirm whether to Save or Not.
Specifies the pipe material to adjust flow data, or choose "Custom" to define the pipe's sound velocity.
Custom
0
Default
UPVC
1
PVC
2
PP
3
PVDF
4
ABS
5
Steel
6
SUS 304
7
SUS 316
8
After selecting the command, please confirm whether to Save or Not.
When A08 is set to "Custom," manually configure the pipe's sound velocity. The unit is determined by B04.
Data Type
float32
Default
2300
Minimum
500
Maximum
10000
After selecting the command, please confirm whether to Save or Not.
Select the type of fluid to adjust the data, or choose "Custom" to define fluid parameters.
Custom
0
Default
Water 20℃
1
Water 50℃
2
Water 75℃
3
Water 100℃
4
Water 125℃
5
Water 150℃
6
Water 175℃
7
Water 200℃
8
Water 225℃
9
Acetone
10
Carbinol
11
Ethanol
12
Alcohol
13
Acetaldehyde
14
Arachis Oil
15
Glycerin
16
Gasoline
17
Benzene
18
Ethyl Benzene
19
Toluene
20
Chloromethane
21
Petroleum
22
Pine Oil
23
Cloroethylene
24
Ketone
25
Glycol
26
Sea Water
27
After selecting the command, please confirm whether to Save or Not.
When A10 is set to "Custom," the fluid sound speed can be manually configured. The unit is determined by B04.
Data Type
float32
Default
1482
Minimum
100
Maximum
3000
After selecting the command, please confirm whether to Save or Not.
When A10 is set to "Custom," the kinematic viscosity of the fluid can be manually configured. The unit is in centistokes (cST).
Data Type
float32
Default
1
Minimum
0.05
Maximum
20000
After selecting the command, please confirm whether to Save or Not.
Reduces data fluctuation. The minimum value is 0 (off), and the default is set to 3.
Data Type
int
Default
3
Minimum
0
Maximum
600
After selecting the command, please confirm whether to Save or Not.
Allows for reversing the direction of fluid flow. The default setting is normal flow.
Normal
0
Default
Inverse
1
After selecting the command, please confirm whether to Save or Not.
Indicates whether negative flow is displayed. The default setting is enabled.
OFF
0
ON
1
Default
After selecting the command, please confirm whether to Save or Not.
Displays a value of zero for flow rates below a specified absolute threshold. The default is set to 0 (disabled), with a maximum threshold of 150,000,000.
Data Type
float32
Default
0
Minimum
0
Maximum
150,000,000
After selecting the command, please confirm whether to Save or Not.
Increases the display update rate to within 0.5 seconds. The default setting is disabled.
OFF
0
Default
ON
1
After selecting the command, please confirm whether to Save or Not.
Eliminates unstable measurement noise. The default setting is enabled, but it can be disabled if required.
Off
0
Low
1
Default
High
2
After selecting the command, please confirm whether to Save or Not.
Activating this function adjusts the signal state based on the environmental conditions. It is recommended to run this function after each reinstallation. This function is identical to Z03.
Ensure the pipe is fully installed and filled. Activating this function corrects zero-flow errors. This function is synchronised with Z04.
Manually adjust or reset the zero offset data to zero.
Data Type
float32
Default
0
Minimum
-100
Maximum
100
After selecting the command, please confirm whether to Save or Not.
After activation, there is a 5-second window to confirm whether to use the A01 Auto Pipe Specification function. The default setting is enabled.
OFF
0
ON
1
Default
After selecting the command, please confirm whether to Save or Not.
Enables or disables the off-pipe state. The default setting is disabled.
OFF
0
Default
ON
1
After selecting the command, please confirm whether to Save or Not.
Enables or disables Beta mode. The default setting is disabled.
Off
0
Default
Low
1
High
2
After selecting the command, please confirm whether to Save or Not.
When A01 is set to "Custom," the pipe outer diameter can be manually configured. The unit is determined by B06. Adjustments will automatically recalculate A04 and A06. A03 to A07 follow the model data range.
When A01 is set to "Custom," the pipe circumference can be manually configured. The unit is determined by B06. Adjustments will automatically recalculate A03 and A06. A03 to A07 follow the model data range.
When A01 is set to "Custom," the pipe thickness can be manually configured. The unit is determined by B06. Adjustments will automatically recalculate A06. A03 to A07 follow the model data range.
When A01 is set to "Custom," the pipe inner diameter can be manually configured. The unit is determined by A02. A03 to A07 follow the model data range.