E Category Options and Descriptions

E01 NC/NO (Normally Closed or Normally Open) Defines whether the circuit is in a conductive (open) or non-conductive (closed) state. Default: Non-conductive.

Options
RS-485 Numbers
Remarks

Normal Close

0

Default

Normal Open

1

After selecting a command, confirm whether to Save or Not Save.

E02 OCT Info. (Output Control Terminal Definition) Configures the operating mode of the OCT. Available options:

  • OFF Alarm AAlarm A + BAlarm A + B + C Flow Alarm (adjustable via D05, D06, D07)

  • Total Alarm (adjustable via D08, D09, D10)

  • Pulse V2 (adjustable via E03, E14; see OCT Output page for usage details).

Default: OFF.

Options
RS-485 Numbers
Remarks

Alarm a

0

Alarm a b

1

Alarm a b c

2

Over Flowrate

3

Over Total

4

Pulse 2.0

5

Off

6

Default

Dosing

7

After selecting a command, confirm whether to Save or Not Save.

E03 Pulse Width Defines the pulse width when E02 is set to Pulse V2. Selectable values: 0.5 ms, 1 ms, 10 ms, 50 ms, 100 ms. Default: 0.5 ms.

Options
RS-485 Numbers
Remarks

0.5 ms

0

Default

1 ms

1

10 ms

2

50 ms

3

100 ms

4

After selecting a command, confirm whether to Save or Not Save.

E04 Modbus ID Configures the transmission protocol for Modbus ID. Default: 1.

Parameter Input Range
Values

Data Type

Int

Default

1

Minimum

1

Maximum

255

E05 Baud (Baud Rate) Defines the Modbus communication baud rate. Options: 9600, 19200, 38400, 57600. Default: Disabled.

Options
RS-485 Numbers
Remarks

9600

19200

38400

57600

OFF

Default

The baud rate cannot be modified via RS485.

After selecting a command, confirm whether to Save or Not Save.

E06 Iout Def. (Analogue Output Definition) Defines the data output via analogue signal. Options: Off, Flow, Velocity, or Cumulative Flow. The unit corresponds to settings in the B category. Default: Off.

Options
RS-485 Numbers
Remarks

Flowrate

0

Velocity

1

Total

2

OFF

3

Default

After selecting a command, confirm whether to Save or Not Save.

E07 4mA Value / 0mA Value Sets the value corresponding to 0 mA or 4 mA. Default: 0. Positive and negative values are accepted, but this value must be less than the 20mA Value.

Parameter Input Range
Values

Data Type

float32

Default

0

Minimum

-100000

Maximum

100000

E08 20mA Value Sets the value corresponding to 20 mA. Default: 100. Positive and negative values are accepted, but this value must be greater than the 4mA Value.

Parameter Input Range
Values

Data Type

float32

Default

100

Minimum

-100000

Maximum

100000

E09 Iout Error (Analogue Output Error Value) Configures the output value when the flow value is outside the range defined by E07 and E08. Options:

  1. 2.4 & 21.6 (Outputs 2.4 mA for values below E07, and 21.6 mA for values above E08).

  2. 2.4 mA.

  3. 21.6 mA.

  4. 0 or 4 & 20 mA (Outputs 4 mA or 0 mA—depending on E13—for values ≤ E07 and 20 mA for values ≥ E08).

Default: 0 or 4 & 20 mA.

Options
RS-485 Numbers
Remarks

2.4 mA & 21.6 mA

0

2.4 mA

1

21.6 mA

2

4mA & 20mA

3

Default

After selecting a command, confirm whether to Save or Not Save.

E11 4mA Adj. uA / 0mA Adj. uA Allows fine-tuning of the 0 mA or 4 mA output value, in units of microamps (uA). Adjustments immediately affect output current for real-time verification.

  • Positive values increase the output. Negative values decrease the output.

Parameter Input Range
Values

Data Type

float32

Default

0

Minimum

-2000

Maximum

2000

E12 20mA Adj. uA Allows fine-tuning of the 20 mA output value, in units of microamps (uA). Adjustments immediately affect output current for real-time verification.

  • Positive values increase the output. Negative values decrease the output.

Parameter Input Range
Values

Data Type

float32

Default

0

Minimum

-2000

Maximum

2000

E13 Analogue Output Range Defines the analogue signal output range. Options:

  • 4–20 mA

  • 0–20 mA

Default: 4–20 mA. Modifying this parameter affects the options in E09. After changes, confirm that E09 settings meet requirements.

Options
RS-485 Numbers
Remarks

4 - 20 mA

0

Default

0 - 20 mA

1

After selecting a command, confirm whether to Save or Not Save.

E14 Pulse V2 Unit Specifies the flow volume per pulse. The unit is determined by B01 settings:

  • L or : Litres (L)

  • U.gal (US Gallon) or ft³: US Gallons

  • I.gal (Imperial Gallon): Imperial Gallons

    Options
    RS-485 Numbers
    Remarks

    0.001 unit

    0

    Default

    0.002 unit

    1

    0.003 unit

    2

    0.005 unit

    3

    0.007 unit

    4

    0.01 unit

    5

    0.02 unit

    6

    0.03 unit

    7

    0.05 unit

    8

    0.07 unit

    9

    0.1 unit

    10

    0.2 unit

    11

    0.3 unit

    12

    0.5 unit

    13

    0.7 unit

    14

    1 unit

    15

    5 unit

    16

    10 unit

    17

    50 unit

    18

    100 unit

    19

After selecting a command, confirm whether to Save or Not Save.

E15 Byte Format Refer to RS485 and Modbus RTU settings for details. This parameter cannot be modified via RS485.

Options
Remarks

N81

Default

N81

O81

E81

After selecting a command, confirm whether to Save or Not Save.

E16 Bit Order Refer to RS485 and Modbus RTU settings for details. This parameter cannot be modified via RS485

Options
Remarks

LSB first

Default

MSB first

After selecting a command, confirm whether to Save or Not Save.

E17 Endianness Refer to RS485 and Modbus RTU settings for details.

Options
RS-485 Numbers
Remarks

Big Endian

0

Default

Little Endian

1

Mid Big Endian

2

Mid Little Endian

3

After selecting a command, confirm whether to Save or Not Save.