A04 Pipe Spec

Definition

Set the pipe outer diameter. Default values vary depending on the selected pipe unit.

DN8
DN10
DN15
DN20
DN25

Default

mm

12

18

22

26

32

inch

0.4724

0.7086

0.8861

1.0236

1.2598

Minimum

mm

9

15

15

18

24

inch

0.3543

0.5905

0.5905

0.7086

0.9448

Maximum

mm

15

22

30

35

43

inch

0.5905

0.8661

1.1811

1.3779

1.6929

circle-info
  1. Millimeters (mm) are used as the base unit, while inches are provided as converted values (conversion reference: 1 inch = 25.4 mm).

  2. If a value exceeds four decimal places, it will be truncated to four decimal places without rounding.

Notes

  1. Pipe size units are determined by B06 (Pipe Size Unit).

  2. Each time the pipe size is changed, please re-execute Z03 Auto Initialize and Z04 Auto Zero Offset to improve measurement accuracy and ensure stable product performance.

  3. On certain firmware versions, a pipe material confirmation page will appear after setting A04 Pipe Wall Thickness. Please select the matching pipe material or choose Custom. If a predefined pipe material is selected, A08 (Pipe Material) will be synchronized automatically.

Operation Procedure

chevron-rightButton Functions (Click to Expand)hashtag

The main unit has four buttons. The bottom line (Line 3) of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:

  • Blue Button - Leftmost button: Return, exit, or save settings.

Gray Arrow Buttons:

  • Left Gray Button: Page up, increase value, or select item

  • Right Gray Button: Page down, decrease value, or select item

  • Orange Button: Rightmost button: Enter settings, edit, or confirm selection

triangle-exclamation
circle-info

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

A04 Options, RS485 Values, Read/Write Commands, and Formats

Input Range of This Parameter
Corresponding Values
RS-485 Value

Custom

Refer to the table above

-

Minimum

-

Maximum

-

Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float 32 data format.