# A22 Auto Zero Offset

### Definition

Recalibrates the flowmeter zero point, resetting the output to zero. Recommended for ultra-low flow measurement to correct zero-flow offset errors and improve measurement accuracy.&#x20;

Starts only after pressing **Start** and automatically updates the **Z04 (Auto Zero Flow)** settings.

It is recommended to perform **Z04** when measuring ultra-low flow (below 10% of the maximum range), as small zero-point offsets can cause significant errors at low flow.

**Example:** With a zero-point offset of 0.2 L/min:

* At an actual flow of 100 L/min, the error is approximately 0.2%.
* At an actual flow of 0.5 L/min, the error can reach nearly 40%.

{% hint style="info" %}
Currently, **A22** does not support Modbus read/write command operation.
{% endhint %}

### Notes

1. Perform **A22**  when the pipeline is <mark style="color:red;">**fully filled with liquid and the liquid is completely stationary**</mark><mark style="color:red;">.</mark> Performing Z04 when the pipe is not full may cause flow measurement errors and unstable readings; if this occurs, repeat Z04 under full-pipe conditions.
2. <mark style="color:red;">**Repeat this procedure after every reinstallation or when changing pipe size or material.**</mark>
3. The **A22** function is equivalent to **Z04**; users can choose either.\
   (Z-series parameters are designed for quick setup, consolidating multiple essential settings to streamline installation.)

### Operation Procedure

<details>

<summary><strong>Button Functions</strong> <em>(Click to Expand)</em></summary>

The main unit has **four buttons**. The **bottom line (Line 3)** of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:

* <img src="/files/bdd75120acea6361d128fd3deec06f1a8277e667" alt="" data-size="line"> **Blue Button -** Leftmost button: Return, exit, or save settings.

Gray Arrow Buttons:

* <img src="/files/2e37e572251371c6020ab6d9e11c1fcdd5ef0369" alt="" data-size="line"> **Left Gray Button:** Previous Page, decrease value, or select item
* <img src="/files/50a67ff5744401e217d5b9e838da3f7909be732f" alt="" data-size="line"> **Right Gray Button:** Next Page, increase value, or select item
* <img src="/files/6c0b22a2048cf9b2afd86fac7a65e414feb967d4" alt="" data-size="line"> **Orange Button:** Rightmost button: Enter settings, edit, or confirm selection

{% hint style="danger" %}
**Do not hold buttons for extended periods to avoid damage.**
{% endhint %}

{% hint style="info" %}
Button functions may change depending on the operation context to improve usability or prevent accidental presses.
{% endhint %}

</details>

```
A22 Automatic Zero Offset Setting Steps
• From the main screen (green backlight), 
press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select  Z. Easy Setup , press Orange Button (Enter).
• Use Gray Arrow Buttons to start Automatic Zero Offset
• Device shows “Start auto Z04” when automatic zero-flow begins.
• After ~30 seconds, it displays “Set zero offset to ____”.
• Then it returns to Z04 Auto Zero Flow, indicating completion.
```

<figure><img src="/files/kqxqYFRsfLafKSEhPaTx" alt="" width="375"><figcaption></figcaption></figure>


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://docs.lorric.com/qr/fu-lt-english-v1/parameters/a-measurement-setup/a22-auto-zero-offset.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.