A15 Damping

Definition

  • A15 Damping: Smooths flow measurement data. Default: 3.

  • Higher value → smoother display, smaller fluctuations, slower response.

  • Lower value → faster response, more sensitive to flow disturbances, larger fluctuations.

Notes

  1. Process Monitoring Application: It is recommended to set A15 higher than the default value to improve display stability and readability, reducing the effect of short-term fluctuations on the displayed value.

  2. Process Control Application: It is recommended to set A15 lower than the default value to minimize measurement delay, allowing flow changes to be reflected in the control system in real time.

  3. Function Description: A15 only affects the smoothness of the display and output; it does not affect the actual measurement accuracy of the flowmeter.

  4. Setting Confirmation: After adjusting A15, it is recommended to observe the system operation for several minutes to verify that the setting meets the application requirements.

  5. Usage Reminder: Setting A15 too high may cause delayed response to flow changes. Adjust according to the actual application scenario.

Operation Procedure

chevron-rightButton Functions (Click to Expand)hashtag

The main unit has four buttons. The bottom line (Line 3) of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:

  • Blue Button - Leftmost button: Return, exit, or save settings.

Gray Arrow Buttons:

  • Left Gray Button: Page up, increase value, or select item

  • Right Gray Button: Page down, decrease value, or select item

  • Orange Button: Rightmost button: Enter settings, edit, or confirm selection

triangle-exclamation
circle-info

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

A15 Damping Setting Steps
• From the main screen(green backlight), 
press Orange Button(Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select A. Measurement, press Orange Button(Enter).
• Use Gray Arrow Buttons to select A15 Damping, press Orange Button(Enter).
• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).
• Press Orange Button(Edit) to enter adjustment mode.
• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
• Press Orange Button(Set) to confirm each digit; repeat for other digits as needed.
• After all digits are set, press Blue Button(Save).
Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).
• Press Blue Button(ESC) to return to the main screen(green backlight).

A15 Options, RS485 Values, Read/Write Commands, and Formats

Input Range of This Parameter
Corresponding Values
RS-485 value

Default

3

-

Minimum

0

-

Maximum

600

-

Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float 32 data format.