E01 NC/NO (Normally Closed or Normally Open)
Defines whether the circuit is in a conductive (open) or non-conductive (closed) state. Default: Non-conductive.
Options
RS-485 Numbers
Remarks
After selecting a command, confirm whether to Save or Not Save.
E02 OCT Info. (Output Control Terminal Definition)
Configures the operating mode of the OCT. Available options:
OFF 、Alarm A、Alarm A + B、Alarm A + B + C Flow Alarm (adjustable via D05, D06, D07)
Total Alarm (adjustable via D08, D09, D10)
Default: OFF.
Options
RS-485 Numbers
Remarks
After selecting a command, confirm whether to Save or Not Save.
E03 Pulse Width
Defines the pulse width when E02 is set to Pulse V2.
Selectable values: 0.5 ms, 1 ms, 10 ms, 50 ms, 100 ms. Default: 0.5 ms.
Options
RS-485 Numbers
Remarks
After selecting a command, confirm whether to Save or Not Save.
Configures the transmission protocol for Modbus ID. Default: 1.
Parameter Input Range
Values
E05 Baud (Baud Rate)
Defines the Modbus communication baud rate. Options: 9600, 19200, 38400, 57600. Default: Disabled.
Options
RS-485 Numbers
Remarks
The baud rate cannot be modified via RS485.
After selecting a command, confirm whether to Save or Not Save.
E06 Iout Def. (Analogue Output Definition)
Defines the data output via analogue signal. Options: Off, Flow, Velocity, or Cumulative Flow. The unit corresponds to settings in the B category. Default: Off.
Options
RS-485 Numbers
Remarks
After selecting a command, confirm whether to Save or Not Save.
E07 4mA Value / 0mA Value
Sets the value corresponding to 0 mA or 4 mA. Default: 0. Positive and negative values are accepted, but this value must be less than the 20mA Value.
Parameter Input Range
Values
Sets the value corresponding to 20 mA. Default: 100. Positive and negative values are accepted, but this value must be greater than the 4mA Value.
Parameter Input Range
Values
E09 Iout Error (Analogue Output Error Value)
Configures the output value when the flow value is outside the range defined by E07 and E08. Options:
2.4 & 21.6 (Outputs 2.4 mA for values below E07, and 21.6 mA for values above E08).
0 or 4 & 20 mA (Outputs 4 mA or 0 mA—depending on E13—for values ≤ E07 and 20 mA for values ≥ E08).
Default: 0 or 4 & 20 mA.
Options
RS-485 Numbers
Remarks
After selecting a command, confirm whether to Save or Not Save.
E11 4mA Adj. uA / 0mA Adj. uA
Allows fine-tuning of the 0 mA or 4 mA output value, in units of microamps (uA). Adjustments immediately affect output current for real-time verification.
Positive values increase the output. Negative values decrease the output.
Parameter Input Range
Values
E12 20mA Adj. uA
Allows fine-tuning of the 20 mA output value, in units of microamps (uA). Adjustments immediately affect output current for real-time verification.
Positive values increase the output. Negative values decrease the output.
Parameter Input Range
Values
E13 Analogue Output Range
Defines the analogue signal output range. Options:
Default: 4–20 mA.
Modifying this parameter affects the options in E09. After changes, confirm that E09 settings meet requirements.
Options
RS-485 Numbers
Remarks
After selecting a command, confirm whether to Save or Not Save.
E14 Pulse V2 Unit
Specifies the flow volume per pulse. The unit is determined by B01 settings:
U.gal (US Gallon) or ft³: US Gallons
I.gal (Imperial Gallon): Imperial Gallons
Options
RS-485 Numbers
Remarks
After selecting a command, confirm whether to Save or Not Save.
Refer to RS485 and Modbus RTU settings for details.
This parameter cannot be modified via RS485.
After selecting a command, confirm whether to Save or Not Save.
Refer to RS485 and Modbus RTU settings for details.
This parameter cannot be modified via RS485
After selecting a command, confirm whether to Save or Not Save.
Refer to RS485 and Modbus RTU settings for details.
Options
RS-485 Numbers
Remarks
After selecting a command, confirm whether to Save or Not Save.