B01 Vol. Unit (Volume Unit)
The unit for instantaneous flow volume, default is L.
Metric units: L (litres), m³ (cubic metres)
Imperial units: U.gal (US Gallon), I.gal (Imperial Gallon), ft³ (cubic feet)
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not Save.
The time unit for flow rate, available options: seconds, minutes, hours, days.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not Save.
B03 Total Unit (Total Volume Unit)
The unit for cumulative and historical volume measurements, default is L.
Metric units: L (litres), m³ (cubic metres)
Imperial units: U.gal (US Gallon), I.gal (Imperial Gallon), ft³ (cubic feet)
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not Save.
B04 Velocity Unit
The unit for flow velocity and sound velocity, default is m/s.
Metric unit: m/s (metres per second)
Imperial unit: ft/s (feet per second)
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not Save.
B05 Dec. Place (Decimal Place)
Specifies the number of decimal places displayed on the LCD/LED screen. Data is truncated without rounding.
Maximum of 4 decimal places, default is 4.
This parameter affects display only, not calculations; cumulative flow uses original data.
After selecting the command, please confirm whether to Save or Not Save.
B06 Pipe Geo. Unit (Pipe Geometry Unit)
The unit for pipe dimensions, synchronised with A03.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not Save.