A01 Auto Pipe Spec: Automatic Pipe Specification Detection
Automatically detects pipe specifications and, upon completion, automatically switches to A02 to display the results. This function is identical to Z02.
A02 Preset Pipe Spec: Pipe Specification
Select a pipe model to adjust the data, or choose "Custom" to define pipe dimensions and material. Supported pipe specifications are listed below:
A02 Different Model Preset Pipe Specification Table. For more details on pipe specifications, refer to ->
Pipe Material Options
RS-485 對應數字
Remarks
After selecting the command, please confirm whether to Save or Not.
A03 Pipe Geo. Unit: Pipe Dimension Unit
Specifies the unit for pipe dimensions, synchronized with B06.
A03 to A07 follow the model data range. For more details on pipe specifications, refer to ->
After selecting the command, please confirm whether to Save or Not.
A04 Pipe OD (Outer Diameter): Pipe Outer Diameter
When A01 is set to "Custom," the pipe outer diameter can be manually configured. The unit is determined by B06. Adjustments will automatically recalculate A04 and A06.
A03 to A07 follow the model data range. For more details on pipe specifications, refer to ->
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A05 Pipe Perimeter: Pipe Circumference
When A01 is set to "Custom," the pipe circumference can be manually configured. The unit is determined by B06. Adjustments will automatically recalculate A03 and A06.
A03 to A07 follow the model data range. For more details on pipe specifications, refer to ->
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A06 Pipe Thickness: Pipe Wall Thickness
When A01 is set to "Custom," the pipe thickness can be manually configured. The unit is determined by B06. Adjustments will automatically recalculate A06.
A03 to A07 follow the model data range. For more details on pipe specifications, refer to ->
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A07 Pipe ID (Inner Diameter): Pipe Inner Diameter
When A01 is set to "Custom," the pipe inner diameter can be manually configured. The unit is determined by A02.
A03 to A07 follow the model data range. For more details on pipe specifications, refer to ->
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A08 Pipe Material: Pipe Material
Specifies the pipe material to adjust flow data, or choose "Custom" to define the pipe's sound velocity.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.
A09 Pipe Sound Speed: Pipe Sound Velocity
When A08 is set to "Custom," manually configure the pipe's sound velocity. The unit is determined by B04.
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A10 Liquid Type: Fluid Type
Select the type of fluid to adjust the data, or choose "Custom" to define fluid parameters.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.
A12 Liquid SS (Sound Speed): Fluid Sound Velocity
When A10 is set to "Custom," the fluid sound speed can be manually configured. The unit is determined by B04.
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A13 K. Viscosity cST: Kinematic Viscosity in cST
When A10 is set to "Custom," the kinematic viscosity of the fluid can be manually configured. The unit is in centistokes (cST).
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A15 Damping: Damping Coefficient
Reduces data fluctuation. The minimum value is 0 (off), and the default is set to 3.
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A16 Direction: Flow Direction
Allows for reversing the direction of fluid flow. The default setting is normal flow.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.
A17 Neg Flow (Negative Flow): Negative Flow Display
Indicates whether negative flow is displayed. The default setting is enabled.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.
A18 Low Flow Cutoff: Force Zero
Displays a value of zero for flow rates below a specified absolute threshold. The default is set to 0 (disabled), with a maximum threshold of 150,000,000.
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A19 Fast Update: High-Speed Update
Increases the display update rate to within 0.5 seconds. The default setting is disabled.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.
A20 Anti-Noise: Noise Reduction
Eliminates unstable measurement noise. The default setting is enabled, but it can be disabled if required.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.
A21 Auto Initialization: Automatic Initialisation
Activating this function adjusts the signal state based on the environmental conditions. It is recommended to run this function after each reinstallation. This function is identical to Z03.
A22 Auto Zero Offset: Automatic Zero Flow Adjustment
Ensure the pipe is fully installed and filled. Activating this function corrects zero-flow errors. This function is synchronised with Z04.
A23 Zero Offset Setting
Manually adjust or reset the zero offset data to zero.
Parameter Input Range
Values
After selecting the command, please confirm whether to Save or Not.
A24 Startup Pipe Check: Initial Pipe Check at Startup
After activation, there is a 5-second window to confirm whether to use the A01 Auto Pipe Specification function. The default setting is disabled.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.
A25 Off-Pipe Zero: Off-Pipe Zero Function
Enables or disables the off-pipe state. The default setting is disabled.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.
Enables or disables Beta mode. The default setting is disabled.
Options
RS-485 Numbers
Remarks
After selecting the command, please confirm whether to Save or Not.