Parameters Table

Parameter categorizing table

Serial number

Name

Function

A

Measure

A group is about flow volume measurement

B

Unit

B group is about unit

C

Total & History

C group is about accumulated and history record

D

Alarm & Info (Alarm & Information)

D group is about alarm and notification.

E

Commute

E group adjust OCT 4-20mA and RS485

F

Simulate

F group is about simulation mode for communication and alarm testing.

G

Time

G group is about date and time set-up

H

System

H group is for system parameters, like brightness and reset.

Parameter description

A Measure parameter list

Serial number

Name

Function

A01

Pipe Type

Choose correct pipe type to get correct result.

A02

Scales Factor

Fine tune flow volume in proportion, default value is 1.0.

A03

Anti-Noise

Eliminate measurement noise. Default on.

A04

Damping

Decrease the fluctuation of data, the lowest is 0, default value is 5 seconds.

A05

Direction

Option to exchange flow direction, default normal.

A06

Neg Flow (Negative Flow)

Option to show reverse flow volume, default on.

A07

Ext. Mode (Extension Mode)

Option to measure flow speed under 0.3m/s, default on.

The video about the monitoring screen, setup screen and A parameter setup.

Watch the video or scan the the QR code below.

B Unit parameter list

Serial number

Name

Function

B01

Vol. Unit (Volume Unit)

Unit of instantaneous flow speed, default litre.

Metric: L, m3

Imperial: US G, UK G, ft3

B02

Time Unit

Unit of flow time seconds, minutes, hours, days. Default minutes.

B03

Total Vol.(Total Volume)

Unit of cumulative and history flow volume, default litre.

Metric: L, m3

Imperial: US G, UK G, ft3

B04

Speed

Unit of flow speed, default m/s.

Metric: m/s

Imperial: ft/s

The video about B parameter setup.

Watch the video or scan the the QR code below.

The video about B parameter setup

C Total & History parameter list

Serial number

Name

Function

C01

Total Method

Set cumulative calculation direction. There are are 3 options available: forward flow, reverse flow, and net flow volume. Switching method will set cumulative volume to zero. Default method is forward flow.

C02

Clean Total

Set cumulative flow volume to 0. First select "Yes" then save.

C03

History Function

To turn on/off history recording function. If turned off, the historical records of the past 14 days will not be displayed on the monitoring screen.

C04

Clean History

Set historical flow volume to 0. First select "Yes" then save.

The video about C parameter setup.

Watch the video or scan the the QR code below.

D Alarm & Information parameter list

Serial number

Name

Function

D01

Paddle Ala. (Paddle Alarm)

Alarm - A

Alarm of paddle wheel error. Default on, but can be turned off.

*Please be aware that magnetic fields or residual magnetism may cause interference to the flowmeter which may result in damage if they are used near the installation site. The purchaser must determine the applicability of the product for its desired use and assumes all risks in connection therewith. LORRIC assumes no responsibility or liability for any omissions or errors in connection with the use of its products.

The alarm used to detect paddle damage or loss is intended to assist on-site operators to find out what is causing equipment abnormalities. Because there are a myriad of variables involved with on-site operation and magnetic fields that come from other equipment, the alarm function is liable to fail. Please do not take the alarm function as a single source for monitoring your flow system. Please contact LORRIC for advanced help.

D02

Mag. Ala. (Magnetic Alarm)

Alarm - A

Alarm when interference from an external magnetic field. Default on, but can be turned off.

D03

Battery Alarm

Alarm - C

Alarm when CR2032 battery is going to be out of battery. Default on, but can be turned off. Attention: Take out the battery and wait one minute, the battery alarm will be displayed.

D04

Flow LMT Ala. (Flow Volume Limit Alarm)

Alarm - B

Alarm when instantaneous flow is out of range. Default on, but can be turned off.

D05

Up Lmt. (Upper Limit)

Define D04 upper limit of normal flow range.

D06

Low Lmt. (Lower Limit)

Define D04 lower limit of normal flow range.

D07

Hyster. Value (Hysteresis Value)

For setting the hysteresis range of turning off alarm.

The Percentage is the proportion of the range between upper and lower limit.

The range is 0-33%, default 3%.

This parameter needs to be positive value.

D08

Total Ala. (Total Alarm)

Alarm - B

Alarm when cumulative flow is out of range. Default on, but can be turned off.

D09

Total Trig. (Total Alarm Trigger Condition)

Set-up the target cumulative value alarm condition. Option "greater than" target or "less than" target . Default "greater than" target.

D10

Total Target

Set up alarm target value of cumulative flow speed.

D11

Iout Alarm

Alarm B

Alarm when it exceeds 4-20mA range. Default on, but can be turned off.

D12

Speed Alarm

Alarm B

Alarm when flow speed exceeds 10m/s. Default on, but can be turned off.

D13

PulseOverlap

Alarm B

Alarm for Pulse Output signal overlaps caused by too long pulse width time parameter E03. Default on, but can be turned off.

The video about D parameter setup.

Watch the video or scan the the QR code below.

E Commute parameter list

Serial number

Name

Function

E01

NC/NO

Under normal conditions(when the oct condition does not meet). OCT should be closed(connected) or open (disconnected). The default is open.

E02

OCT Info

There are options below are options to define OCT output, default Alarm A:

Alarm A,

Alarm A+B,

Alarm A+B+C,

flow volume alarm(adjusted by D04 D05 D06),

cumulation alarm(adjusted by D08 D09 D10),

paddle wheel pulse output( adjusted by E03).

E03

Pulse Width

If paddlewheel pulse is selected in E02, you set the pulse width to 0.5 ms (millisecond), 1ms, or 10ms. The default is 0.5ms. Every 180° paddlewheel rotation will output a pulse. Please refer to the Pulse- flow volume conversion chart for flow volume calculation.

E04

Modbus Addr (Modbus Address)

Set up RS485 location, default 1.

E05

Baud

You can set-up RS485 communication baud rate at 9600 or 19200. The default is 9600. After version FW v1.0.3 options 38400 or 57600 are available.

E06

Iout Def. (Iout Definition)

For setting output data of 4-20mA, 2 options are available, flowrate and velocity. The unit is according to parameter group B. Cumulative flow volume are available after FW version v1.0.3.

E07

4mA Value

Set up corresponding value to 4mA, default 0. Both positive and negative values work. The value should be smaller than 20mA.

E08

20mA Value

Set up corresponding value to 20mA, default 100. Both positive and negative values work. The value should be larger than 4mA.

E09

Iout Error

Set up the outputted value when the flow volume is not between E07 and E08.

3 Options modes:

  1. 2.4 and 21.6 mA (When less than E07, it outputts 2.4mA. When greater than E08, it outputs 21.6mA)

  2. 2.4 mA

  3. 21.6 mA

Default 2.4 & 21.6 mA.

E11

4mA Adj. uA (4mA Output Adjustment in microAmp)

Fine tune 4mA outputted value, the unit is uA.

After fine tuning and saving the setting, output current will be varied for immediate check.

Positive value increases output current, negative value decreases output current.

E12

20mA Adj. uA (20mA Output Adjustment in microAmp)

Fine tune 20mA outputted value, the unit is uA.

After fine tuning and saving the setting, ooutput current will be varied for immediate check.

Positive value increases output current, negative value decreasess output current.

The video about E parameter setup.

Watch the video or scan the the QR code below.

F Simulate parameter list

Serial number

Name

Function

F01

Sim. Volume (Simulation Volume)

Set up simulate value of instantaneous flow speed, the unit will be the one set in parameter group B.

F02

Sim. Mode (Simulation Mode)

Simulation function will be activated after selecting “ on ”, default off.

OCT (besides pulse), RS485 and 4-20mA will be operated according to the changes of simulated data and cumulative flow.

Simulate mode will be reminded on the screen.

Please be noted that cumulative flow and history data will be affected.

F03

Sim. PulseHz (Simulation PulseHz)

Under this page, pulse signal will be outputted according to set frequency within value 0 to 300. Please set up NC/NO of pulse according to parameter E01, and pulse width according to parameter E03.

The video about F parameter setup.

Watch the video or scan the the QR code below.

G Time parameter list

Serial number

Name

Function

G01

Set Year

Set up yaer

G02

Set Month

Set up month

G03

Set Day

Set up day

G04

Set Time

Set up time, accurate to minute.

G05

Date Format

Set up format of date display on screen: YYYY-MM-DD, YYYY/MM/DD,DD/MM/YYYY, MM/DD YY

Default YYYY-MM-DD

The video about G parameter setup.

Watch the video or scan the the QR code below.

H System parameter list

Serial number

Name

Function

H01

LED Bri. (LED Brightness)

Set up brightness of LED with 3 levels, default middle level.

H02

LCD Bri. (LCD Brightness)

Set up brightness of LCD with 3 levels, default middle level.

H03

Green Light

Set up time of LCD green lighting with 3 durations, 10s, 60s, and always on.

H04

Language

Set up language. Default english and traditional chinese (Provided after ver. FW v1.1.0)

H05

Paddle Init. (Paddle Initialization)

Press "Start" to initialize the paddle. User needs to rotate the paddle at least 20 rounds until the screen shows "Init. Succeed". If screens shows "Init. Failed", please try again. If the problem persists, please replace your paddle.

H06

Reset

Clear all the data (include date and cumulative data), reset all parameters to default values.

The video about H parameter setup.

Watch the video or scan the the QR code below.

Last updated