Mechanical and electrical installation

Definition of M12 8Pin connector

  1. Connect the cable inside the box to the power supply and communication system. The unused color cables should be covered or protected to avoid electric leakage or short circuit.

  2. Before removing the top housing of the device and wiring, the device should be powered off and the M12 8pin cable should be removed.

  3. Due to space constraints, the sticker on the side of the device and subsequent illustrations only use cable contact definitions.

  4. Please see below images to know where each cable is connected.

Mechanical and electrical spec of device

Wiring diagram of DC power supply

Wiring diagram of 4-20mA analog output

This function will only work with power supplies through the brown and blue lines. The following diagram is not showing power supply wiring for readability.

Below is a complete wiring diagram.

4-20mA negative terminal can not be connected electrically with DC power negative terminal which means it should work with independent power supply. However, wrong cable connecting may cause functional damage.

OCT NPN output wiring diagram

This function will only work with power supplies through the brown and blue lines. The following diagram is not showing power supply wiring for readability.

OCT PNP output wiring diagram

This function will only work with power supplies through the brown and blue lines. The following diagram is not showing power supply wiring for readability.

RS485 2 wire wiring diagram

This function will only work with power supplies through the brown and blue lines. The following diagram is not showing power supply wiring for readability.

Device contact definitions

  1. Due to space constraints, the sticker on the side of the device only use cable contact definitions.

  2. The number of the device contact and the cable contact is mirrored and symmetrical.

  3. Please see below device contact number:

How to replace a date battery of FP-AS510

Applicable to all the models from HW v1.0.0 to v1.3.1

1.Take off top housing and then remove drained battery by the plastic ruler or other tools so that the device will not be scratched.

2.Put in a new CR2032 battery by the position with negative electrode (-) side towards battery base and positive electrode (+) side towards outside. Insert the battery on the side where there is a cable attached.

3. Press the battery on the other side where there is no cable attached. Finish.

  1. Please turn off the device and unplug before disassembling it.

  2. Take off four hex socket head cap screws on the top housing according to above image.

  3. Seperate top and bottom housing.

    1. Be aware when separating the top and bottom housing because there are many wires connecting both.

  4. Take off the drained battery from the battery base on the side wall of the bottom housing.

  5. Put in a new battery. Positive electrode(+) side should be towards the battery base.

  6. Rotate top housing to desired viewing orientation.

    1. In order to prevent damaging the cables, do not rotate the top housing too many times with unnecessary pulls.

  7. In the bottom housing there is an o-ring with an inner groove for a thread hole . For it to work properly, please confirm there is no particle, fiber or any object that may break the sealing function before attaching the top and bottom housing.

  8. Combine the top with bottom housing.

  9. Tighten four screws on top housing.

    1. Too much torque may damage the device.

  10. Finish.