Device

Exterior

Device Functionalities

LED monitor mainly provides high contrast numbers display for high visibility even in a dark environment.

LED monitor also provides 5 displayed digits within a decimal point which mainly displays real-time and accumulated flow volume data. For more digits of data, such as historical data and long-term cumulative data, please see the LCD.

LCD monitor allows display for 16 English characters in 3 rows . The first row is current content with units. The second row is data. The third row is a definition of each button.

The LCD screen has a 3-color backlit display. Definition of each color light as below:

  1. Geen: Real-time data, device is working normally.

  2. Red:

    1. Blinking red light: Alarm A- Major error. Please see the “Alarm and Description of System Conditions” page for more details.

    2. Continuous red light: Alarm B- Current status of the device meets the parameters set-up by the user so must pay attention to it. Alarm C- The calendar battery needs to be replaced.

  3. Orange: Non real-time data display. Historical data query and set-up page display.

There are 4 buttons on the device, the definition of each button will be displayed at the upper row on the LCD screen. Generally, the leftmost button stands for “Exit” or “Back to last page”. The middle 2 buttons are for users to review page by page, modify data, or switch option. The rightmost button stands for “Set” or “Confirm”. Previously mentioned definitions are only for general use. For some other use scenarios, button definitions are modifiable for usability or error proof.

Package contains

Besides other agreements, our standard package contains:

  • FP-AS510 Device * 1

  • 2m 8pin cable * 1

  • Quick starting manuel * 1

  • Screws to connect the device and paddle set (Locking screws with your convenient side is suggested. It is unnecessary locking screws both sides)

Paddle set and T-connector can be purchased separately

Size, material, model

Parts

Material

Screws for top side housing

Stainless steel 304

Screen cover

PMMA

Whole housing without screen cover

PC+GF

Buttons

PET

2 meters long 8 pin cable with insulation cover

Union nuts of 2 meters long 8 pin cable

Oring of 2 meters long 8 pin cable

Sealing ring for top housing

M12 8 pin

Mechanical and electrical spec of device

Power supply

Power supply

Working specification

12V ~ 36V 90mA (4-20mA and RS485 current are not included.)

Reverse voltage protection

Yes

Overcurrent protection

200mA

OCT output

OCT output

Power supply spec

5V~30V 200mA

Reverse voltage protection

Yes

NPN connection

Yes

PNP connection

Yes

Maximum output frequency

700Hz

Overcurrent protection

200mA

Analog output (4-20mA, 0-20mA)

Analog output (4-20mA, 0-20mA)

Wiring

4 wiring

Power supply

Device is self-powered, no external power supply required.

Maximum output range

0~21.6mA

Response time

<1s, the same as screen data.

Loop Resistance

1100 ohm @ 36V

600 ohm @ 24V

250 ohm @ 12V

Digital communication Modbus RTU RS485

Digital communication Modbus RTU RS485​

Baud

9600, 19200, 38400, 57600

Wiring

2 wiring

Nodes on the bus

32

Last updated