F02 Sim. Mode
Definition
Enables or disables the flow meter’s simulation mode. In simulation mode, the meter displays user-defined virtual flow values instead of actual sensor measurements. Default: Off.
Notes
After enabling this function, set the simulated flow in F01.
OCT (excluding pulses), RS485, and analog current outputs will operate based on simulated data and totalized flow, affecting both totalized values and historical records.
Operation Procedure
Button Functions (Click to Expand)
The main unit has four buttons. The bottom line (Line 3) of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:
Blue Button - Leftmost button: Return, exit, or save settings.
Gray Arrow Buttons:
Left Gray Button: Page up, increase value, or select item
Right Gray Button: Page down, decrease value, or select item
Orange Button: Rightmost button: Enter settings, edit, or confirm selection
Do not hold buttons for extended periods to avoid damage.
Button functions may change depending on the operation context to improve usability or prevent accidental presses.
Simulation Mode Instructions (Click to Expand)
Simulation Mode is used to output simulated flow values for testing communication and signals. Related parameters include F01 Sim. Volume and F02 Sim. Mode, and additional settings depending on communication type:
4–20 mA: E06 Iout Def.
NPN Pulse: E01 NO/NC, E02 OCT Definition
RS485: E04 Modbus ID, E05 Baud Rate
F02 Turning on Simulation ModeF02 Sim. Mode Setting Steps• From the main screen (green backlight), press Orange Button (Setup) to enter settings (orange backlight).• Use Gray Arrow Buttons to select F. Simulate , press Orange Button (Enter).• Use Gray Arrow Buttons to select F02 Sim. Mode, press Orange Button (Enter).• Use Gray Arrow Buttons to choose the options. • Press Orange Button (Set) to confirm.• Press Blue Button (ESC) to return to the main screen (green backlight).F01 Simulated Flow Setting ProcedureF01 Sim. Volume Setting Steps• From the main screen(green backlight), press Orange Button(Setup) to enter settings (orange backlight).• Use Gray Arrow Buttons to select F. Simulate, press Orange Button(Enter).• Use Gray Arrow Buttons to select F01 Sim. Volume, press Orange Button(Enter).• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).• Press Orange Button(Edit) to enter adjustment mode.• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.• Press Orange Button(Set) to confirm each digit; repeat for other digits as needed.• After all digits are set, press Blue Button(Save). Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).• Press Blue Button(ESC) to return to the main screen(green backlight).Communication-Based Parameter SettingsFor 4–20 mA Wiring, Enable E06 Analog Output DefinitionE06 Iout Def. Setting Steps• From the main screen (green backlight), press Orange Button (Setup) to enter settings (orange backlight).• Use Gray Arrow Buttons to select E. Commute , press Orange Button (Enter).• Use Gray Arrow Buttons to select E06 Iout Def., press Orange Button (Enter).• Use Gray Arrow Buttons to choose the options. • Press Orange Button (Set) to confirm.• Press Blue Button (ESC) to return to the main screen (green backlight).For NPN Wiring, Enable E01 NO/NC and E02 OCT DefinitionE01 NC/NO and E02 OCT Setting Steps• From the main screen (green backlight), press Orange Button (Setup) to enter settings (orange backlight).• Use Gray Arrow Buttons to select E. Commute, press Orange Button (Enter).• Use Gray Arrow Buttons to select E01 NC/NO, press Orange Button (Enter).• Choose the options NC/NO, press Orange Button (Set).• The screen returns automatically to E. Commute.• Use Gray Arrow Buttons to select E02 OCT Info, press Orange Button (Enter).• Press Blue Button (ESC) to return to the main screen. (orange backlight indicates Simulation Mode is active)• Verify at the receiving end that the corresponding OCT signal is output correctly.For RS485 Wiring, Enable E04 Modbus ID and E05 Baud Rate.E04 Modbus ID and E05 Baud Rate Setting Steps• From the main screen (green backlight), press Orange Button (Setup) to enter settings (orange backlight).• Use Gray Arrow Buttons to select E. Commute, press Orange Button (Enter).• Use Gray Arrow Buttons to select E04 Modbus ID, press Orange Button (Enter).• Use Gray Arrow Buttons to adjust the Modbus ID, press Orange Button (Set).• The screen returns automatically to E. Commute.• Use Gray Arrow Buttons to select E05 Baud Rate, press Orange Button (Enter).• Choose the options, press Orange Button (Set). (Options: Off, 57600, 38400, 19200, 9600)• Press Blue Button (ESC) to return to the main screen. (orange backlight indicates Simulation Mode is active)• Verify at the receiving end that the corresponding OCT signal is output correctly.F02 Simulation Mode (Turn Off After Testing)If communication is normal, return to F02 Simulation Mode and turn it off. The main screen will return to green backlight.
F02 Options, RS485 Values, Read/Write Commands, and Formats
OFF (Default)
OFF
0
ON
ON
1
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Int16 data format.