FU-TX 310 Clamp-on Ultrasonic Flowmeter Manual v3.
  • This instruction manual version is v3.0
  • Firmware update date and version function description
  • 1. Product Introduction
    • 1.1 Specifications
    • 1.2 Important concepts
    • 1.3 Product features
    • 1.4 Product applications
    • 1.5 Measuring principle
    • 1.6 Product contents
    • 1.7 User interface
      • 1.7.1 Screen interfaces
      • 1.7.2 Buttons
      • 1.7.3 LED message
  • 2. Installation Method
    • 2.1 Power
    • 2.2 Ultrasonic probe
      • 2.2.1 Probe wiring
      • 2.2.2 Meter wiring
      • 2.2.3 Installation location
      • 2.2.4 Installation method
      • 2.2.5 ​Probe installation method
    • 2.3 Temperature sensor/Analog output
      • 2.3.1 Temperature sensor
      • 2.3.2 Analog output
      • 2.3.3 RS485
    • 2.4 Installation check
      • 2.4.1 RSSI (Signal Strength)
      • 2.4.2 Q value
      • 2.4.3 Flight time
  • 3. Output Interface
  • 4. Setting up the Interface
    • 4.1 Intial menu settings
    • 4.2 Units
    • 4.3 Meter (Measurement Settings)
    • 4.4 Communication
    • 4.5 Other
    • 4.6 Save
  • 5. Error Messages and Solutions
  • 6. Other Conditions
  • 7. MODBUS Communication Protocol
  • 8. Cautions
  • 9. Sound Speeds Data
  • LORRIC Integration and Application Center
Powered by GitBook
On this page

Was this helpful?

Export as PDF
  1. 2. Installation Method
  2. 2.4 Installation check

2.4.3 Flight time

The seventh interface of the display output interface displays all values related to ultrasonic signal flight time (Time of Flight, or ToF). TOM (as shown in the picture) represents measured flight time, and TOS the calculated theoretical flight time. Ideally, the two values should be equal. If TOM/TOS yields a value that exceeds 100% ± 3%, the following steps should be taken:

● Check if the relevant setting parameters are correct. Check pipe material, pipe diameter, whether there is a liner, fluid type, and other relevant parameters. Check to ensure that the probe is mounted correctly.

● Check to ensure that the probe mounting distance is correct.

If> 100%, adjust upstream and downstream probes and bring them closer together.

If <100% adjust distance of probe, by moving them further apart.

Previous2.4.2 Q valueNext3. Output Interface