LORRIC Docs For Partners
FU-TX 310 Clamp-on Ultrasonic Flowmeter Manual v2.
FU-TX 310 Clamp-on Ultrasonic Flowmeter Manual v2.
  • This instruction manual version is v2.1
  • 1. Product Introduction
    • 1.1 Specifications
    • 1.2 Important concepts
    • 1.3 Product features
    • 1.4 Product applications
    • 1.5 Measuring principle
    • 1.6 Product contents
    • 1.7 User interface
      • 1.7.1 Screen interfaces
      • 1.7.2 Buttons
      • 1.7.3 LED message
  • 2. Installation Method
    • 2.1 Power
    • 2.2 Ultrasonic probe
      • 2.2.1 Probe wiring
      • 2.2.2 Meter wiring
      • 2.2.3 Installation location
      • 2.2.4 Installation method
      • 2.2.5 ​Probe installation method
    • 2.3 Temperature sensor/Analog output
      • 2.3.1 Temperature sensor
      • 2.3.2 Analog output
      • 2.3.3 Wiring method
    • 2.4 Installation check
      • 2.4.1 RSSI (Signal Strength)
      • 2.4.2 Q value
      • 2.4.3 Flight time
  • 3. Output Interface
  • 4. Setting up the Interface
    • 4.1 Intial menu settings
    • 4.2 Units
    • 4.3 Meter (Measurement Settings)
    • 4.4 Communication
    • 4.5 Other
    • 4.6 Save
  • 5. Error Messages and Solutions
  • 6. Other Conditions
  • 7. MODBUS Communication Protocol
  • 8. Cautions
  • 9. Sound Speeds Data
  • LORRIC Integration and Application Center
Powered by GitBook
On this page

4. Setting up the Interface

Previous3. Output InterfaceNext4.1 Intial menu settings

Last updated 5 years ago

Press the SET​ button in the lower left corner of the output interface to enter the system setup interface, which contains:

■ Initial

■ Unit

■ Meter

■ Wrreless

■ Others

■ Save

The initial setting (Initial​) is the first step. It is essential to complete this step first. Incomplete or incorrect data entry will prevent the device from operating normally. After input has been completed, the Save function must be performed to save the settings, otherwise the set data will be lost after powering off. Initial​ menu settings are as follows:

Choose [1.Initial ] and press [ENT] button to go into the initial setup. Then, press the arrow to choose any of the setup options. Press the [ENT] button to configure the function. Press the [Back] button to return to the previous page.

Choose [2.Unit] and press [ENT] button to go into the initial setup. Then, press the arrow to choose any of the setup options. Press the [ENT] button to configure the function. Press the [Back] button to return to the previous page.

Choose [3.Meter] and press [ENT] button to go into the initial setup. Then, press the arrow to choose any of the setup options. Press the [ENT] button to configure the function. Press the [Back] button to return to the previous page.

Choose [4.Communication] and press [ENT] button to go into the initial setup. Then, press the arrow to choose any of the setup options. Press the [ENT] button to configure the function. Press the [Back] button to return to the previous page.

Choose [5.Others] and press [ENT] button to go into the initial setup. Then, press the arrow to choose any of the setup options. Press the [ENT] button to configure the function. Press the [Back] button to return to the previous page.

To save your setup, you must execute the Save function to save, choose [6.Save] and press [ENT] button. Press [Yes] to confirm your save or press [No] to go back to the [System Setup] page.