1.1 Specifications
■ Installation method | Tube clamps |
■ Measurement principle | Time differential |
■ Flow rate range | ±0.1~20m/s |
■ Measurement accuracy | <±2 % |
■ Response time | <1 second |
■ Resolution | 0.0001 m/s |
■ Wired communication | Analog output 4-20 mA Modbus RS485 Two-line |
■ Probe-to-host distance | 10M |
■ Temperature measurement | Two sets of external PT1000 |
■ Temperature range | -100~300℃ with 0.1°C resolution |
■ Device working temperature | -10~60°C(14~140°F) |
■ Applicable fluid | Clean water, oil or chemical with minor impurities |
■ Wall temperature | Standard probe: 0~80°C High temperature probe: 0~150°C |
■ Power consumption | < 2W |
■ Transient flow | Instantaneous flow, flow velocity, time differential display |
■ Cumulative flow | Positive and negative cumulative, net flow display |
■ Units | Metric or Imperial units |
■ Display | 128 x 64 LCD backlight display |
■ Operation buttons | 4 key touch buttons |
■ Security | Keyboard lock, power-loss data protection |
■ Shell | ABS plastic, 145 x 90 x 45 mm |
■ Power | 9~30VDC 100-240 50/60Hz AC transformer |
■ Applicable pipe material | Cast iron, carbon steel, stainless steel, PVC pipe and other |
■ Applicable pipe diameter (mm) | 50mm~200mm(standard equipment) |
■ Probe waterproof rating | General probe IP61 Glue probe IP65 Waterproof resistance probe IP68 |