1.1 Specifications
Last updated
Last updated
■ Installation method
Tube clamps
■ Measurement principle
Time differential
■ Flow rate range
±0.1~20m/s
■ Measurement accuracy
<±2 %
■ Response time
<1 second
■ Resolution
0.0001 m/s
■ Wired communication
Analog output 4-20 mA
Modbus RS485 Two-line
■ Probe-to-host distance
10M
■ Temperature measurement
Two sets of external PT1000
■ Temperature range
-100~300℃ with 0.1°C resolution
■ Device working temperature
-10~60°C(14~140°F)
■ Applicable fluid
Clean water, oil or chemical with minor impurities
■ Wall temperature
Standard probe: 0~80°C
High temperature probe: 0~150°C
■ Power consumption
< 2W
■ Transient flow
Instantaneous flow, flow velocity, time differential display
■ Cumulative flow
Positive and negative cumulative, net flow display
■ Units
Metric or Imperial units
■ Display
128 x 64 LCD backlight display
■ Operation buttons
4 key touch buttons
■ Security
Keyboard lock, power-loss data protection
■ Shell
ABS plastic, 145 x 90 x 45 mm
■ Power
9~30VDC
100-240 50/60Hz AC transformer
■ Applicable pipe material
Cast iron, carbon steel, stainless steel, PVC pipe and other
■ Applicable pipe diameter (mm)
50mm~200mm(standard equipment)
■ Probe waterproof rating
General probe IP61
Glue probe IP65
Waterproof resistance probe IP68