Alarms and System Status

Alarm A - Critical Failure

The system error level follows the Failed message class defined in NAMUR NE107.

Triggered during flow sensing errors or main unit failures. The main screen LED displays Error, and the LCD backlight flashes red.

Alarm A Error Code Definition: Bits are defined from MSB (Bit 7) to LSB (Bit 0), with each bit representing the following:

Alarm bit
Alarm Definition
Alarm Code
Trigger Condition

7

na

na

na

6

na

na

na

5

na

na

na

4

Flowmeter Not on Pipeline

D04

Flowmeter is not on the pipeline or is not correctly installed on the pipeline

3

IC Failure = 1

D15

Iout (0/4–20 mA) IC Failure = 1. Parameter D14 must be in the "on" or "do not disturb" state

2

IC Overheating = 1

D15

Iout (0/4–20 mA) IC Overheating = 1. Parameter D14 must be in the "on" or "do not disturb" state

1

Abnormal Pipe Diameter Sensing Element

D02

Abnormality in the module used to detect pipe diameter

0

Ultrasonic Sensor Communication Error

D01

Triggered when the system detects abnormal ultrasonic component communication. Typical causes include opening the device, rotating the display cover excessively and straining internal wiring, or external forces affecting ultrasonic communication

Alarm B – Out of Specification

The system error level follows the Out of Specification message class defined in NAMUR NE107.

Used for flow beyond the set range or other conditions requiring attention. Configurable under the D parameter category. The main screen LED display is unchanged; the LCD backlight turns solid red when active.

Alarm B Error Code Definition: Bits are defined from MSB (Bit 7) to LSB (Bit 0), with each bit representing the following:

Alarm-bit
Alarm Definition
Alarm Code
Trigger Condition

7

Dosing Target Reached

D16

The dosing target set at C04 is reached

6

Totalized Flow Exceeds Target

D09

The totalized counter reaches the preset target value at D11, and the accumulation condition at D10 is met

5

Value Exceeds Analog Output Upper Limit

D12

Flow value exceeds the 20 mA upper limit set at E08. For example, if E08 is set to 100 L, the actual flow is 150 L

4

Value Below Analog Output Lower Limit

D12

Flow value falls below the 4 mA lower limit set at E07. For example, if E07 is set to 0 L, the actual flow is −10 L

3

Flow Velocity > 6 m/s or < −6 m/s

D13

Triggered when absolute flow velocity exceeds 6 m/s; cleared when absolute flow velocity returns to 6 m/s or below

2

OCT Timing Too Slow

D14

In pulse output mode, OCT timing is set too slow, preventing correct pulse output

1

Instantaneous Flow Above Upper Limit

D05

Triggered when instantaneous flow exceeds the upper limit set at D06; cleared when flow drops below D06 minus the hysteresis value

0

Instantaneous Flow Below Lower Limit

D05

Triggered when instantaneous flow falls below the lower limit set at D07; cleared when flow rises above D07 plus the hysteresis value

Alarm C - Maintenance Required

The system error level follows the Maintenance Required message class defined in NAMUR NE107.

Used to indicate that the flowmeter requires maintenance. The main screen LED display is unchanged; the LCD backlight turns solid red when active.

Alarm C Error Code Definition: Bits are defined from MSB (Bit 7) to LSB (Bit 0), with each bit representing the following:

Alarm-bit
Alarm Definition
Alarm Code
Trigger Condition

7

na

na

na

6

na

na

na

5

na

na

na

4

na

na

na

3

Poor Ultrasonic Signal

D03

Signal is 0 bars

2

Poor Signal Due to Installation Issue

D04

Poor signal caused by installation issues. Evenly tighten the flowmeter to achieve optimal signal quality

1

Excessive Line Impedance or No Wiring Connection

D15

Iout (0/4–20 mA) excessive line impedance or no wiring connection = 1; normal = 0. Parameter D14 must be in the "on" or "do not disturb" state

0

Battery Low

D09

Alarm D - Function Check

The system error level follows the Function Check message class defined in NAMUR NE107.

When the main unit is being operated manually and is not on a monitoring screen, this bit becomes 1. On the settings screen, the LCD backlight turns solid orange.

Alarm D Error Code Definition: Bits are defined from MSB (Bit 7) to LSB (Bit 0), with each bit representing the following:

Alarm-bit
Alarm Definition
Alarm Code
Trigger Condition

7

na

na

na

6

na

na

na

5

na

na

na

4

na

na

na

3

na

na

na

2

na

na

na

1

na

na

na

0

Main Unit Not on Monitoring Screen

na

Becomes 1 in simulation mode or on Setup screens

System Status

Status codes are defined from MSB (Bit 7) to LSB (Bit 0), with each bit representing the following:

Alarm-bit
Alarm Definition
Alarm Code
Trigger Condition

7

na

na

na

6

na

na

na

5

na

na

na

4

na

na

na

3

na

na

na

2

na

na

na

1

na

na

na

0

OCT Status

na

NC – Normally Closed (conducting; Close = 1); NO – Normally Open (not conducting; Open = 0). Pulse output status is not included