A27 Wedge Method
Definition
Sets the transducer mounting method. The method determines the ultrasonic path through the pipe, and therefore the spacing of the two transducers will be shown in A29 Probe Spacing.
Notes
Set A27 before installing the transducers — the required spacing in A29 depends on it.
After changing A27, re-run Z03 Auto Initialization and Z04 Auto Zero Offset.
Operation Procedure
A27 Wedge Method Setting Steps
• From the main screen (green backlight), press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select A. Measure, press Orange Button (Enter).
• Use Gray Arrow Buttons to select A27 Wedge Method, press Orange Button (Enter).
• Use Gray Arrow Buttons to choose the mounting method.
• Press Orange Button (Set) to confirm.
• Press Blue Button (ESC) to return to the main screen (green backlight).
A27 Options, RS485 Values, Read/Write Commands, and Formats
Options
Description
RS-485 value
Z
Single pass, opposite sides
0
V (Default)
Double pass, same side
1
N
Triple pass, same side
2
W
Quadruple pass, same side
3
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Int16 data format.