A03 Pipe Geo Unit
Definition
This function sets the unit for the flow meter’s tube size. The default unit is mm.
Notes
After executing this function, the setting for B06 (Tube Size Unit) will be automatically updated.
Each time the pipeline size is changed, perform Z03 Auto Initialize and Z04 Auto Zero Offset to ensure accurate flow readings and maintain stable product quality.
Operation Procedure
A03 Pipe Geo Unit Setting Steps
• From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select A. Measurement , press Orange Button (Enter).
• Use Gray Arrow Buttons to select A03 Pipe Geo Unit, press Orange Button (Enter).
• Use Gray Arrow Buttons to choose the options.
• Press Orange Button (Set) to confirm.
• Press Blue Button (ESC) to return to the main screen (green backlight).
A03 Options, RS485 Values, Read/Write Commands, and Formats
Options
Description
RS-485 Value
mm (Default)
mm
0
inch
inch
1
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Int16 data format.