Parameter Settings
Setup Procedure
Power on the device as wired in [Electrical Installation].

Once powered, the display shows the main screen (green backlight) — the starting point for every step below.
Language Settings
The default language is English. The unit supports English, Traditional Chinese, and Simplified Chinese. To change the language, go to H04 Language.

Pipe Setting
Complete settings in order: A01, A02 A04, A06, A08.
A01 DN Pipe Choose
Select the nominal pipe diameter (DN). This is the first pipe-setup parameter — the DN chosen here determines which specs are available in A02 Preset Pipe Spec. Default: DN 65 (2½").

A02 Preset Pipe Spec
This function is used to select the piping specifications used by the flow meter, with CNS PVCU as the default.

A04 Pipe Outer Diameter
Set the pipe outer diameter by adjusting each digit individually. Press Left Gray Button to increase a digit and Right Gray Button to decrease it. After setting a digit, press Orange Button (Set) to confirm before adjusting the next digit. Please refer to A04 Pipe Outer Diameter.

A06 Pipe Thickness
Set the pipe wall thickness. Each digit can be adjusted individually. Use Left Gray Button to increase and Right Gray Button to decrease each digit. Please refer to A06 Pipe thickness

A08 Pipe Material
Set the pipe material. If no matching option is available, select custom. Please refer to A08 Pipe Material.

custom(default) 、PE、PFA、Copper、SUS316、SUS304、Steel、ABS、PVDF、PP、PVC、PVCU
Probe Installation Method Setting
A27 Wedge Method
Set the pipe wedge method for the probe. For how to determine the wedge method, please refer to A27 Wedge Method.

Check Probe Spacing for Installation
Once Pipe Settings and Wedge Method (Steps 2–3) are entered, the device calculates the required spacing between the probes automatically. This value updates automatically whenever Pipe Settings or Wedge Method change. Record the spacing value and use it when installing the probes — see Mechanical Installation, "Mount the TM-1 Transducer on the Guided Track Clamp."

Fluid Sound Velocity Setting
After completing the pipe and wedge method settings, ensure the pipe is fully filled and run Z03 Auto Initialize. The flowmeter will automatically detect the ultrasonic sound velocity of the measured fluid and configure the operating parameters accordingly. Refer to: Z03 Auto Initialize.


Main Screen Status Check
After setup, return to the main screen. If settings are correct, the flow rate will be displayed. Check the flow signal indicator at the lower-left corner:
1 bar: Poor signal quality
< 1 bar: Possible air bubbles or interference; "Noisy" may appear
If the signal is low, adjust the sensor installation according to the mounting guidelines and check whether the signal improves.

Troubleshooting
If any of the following issues occur during installation or setup, adjust the system as instructed.
1. Rapid Fluctuating Readings
If flow readings fluctuate rapidly, adjust A15 Damping to reduce data swings. The default damping value is 3; higher values reduce fluctuations. Set according to desired measurement stability and response speed. Refer to A15 Damping.
2. No Zero at Zero Flow
Go to A18 Low Flow Cutoff to configure forced zero.
Purpose:
Prevents false flow readings caused by signal drift at zero flow.
Parameter Description:
A18 = 0 (default): Shows all very low flows; suitable for applications requiring detection of tiny flow.
A18 = X (user-defined): Forces the reading to zero when flow < X.
Recommended Setting:
For non-low-flow applications, set an appropriate cutoff value to improve zero stability and prevent flow display during stoppage.
Refer to model-specific lower limit suggestions below.
Refer to A18 Low Flow Cutoff.
3. Incorrect Positive/Negative Sign Display
If the flowmeter shows the wrong sign, adjust the detection direction in A16 Direction. Refer to A16 Direction.
4. Unit Adjustment
To change the display unit, go to B Unit. Refer to B Unit.
Advanced Calibration Settings
Refer to the full parameter table for detailed settings. See: Full Parameter Reference.