Parameter Settings
Setup Procedure
1
Language Settings
H04 Language Setting Steps
- From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
- Use Gray Arrow Buttons to select H.System, press Orange Button (Enter).
- Select H04 Language, press Orange Button (Enter).
- Choose language, press Orange Button (Set) to confirm.
- Press Blue Button (ESC) to return to main screen.
2
Pipe Setting
A04 Pipe Outer Diameter
A04 Pipe Outer Diameter Setting Steps
- From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
- Use Gray Arrow Buttons to select A.Measurement, press Orange Button (Enter).
- Use Gray Arrow Buttons to select A04 Pipe Outer Diameter,
press Orange Button (Enter).
- Use Gray Arrow Buttons to choose the digit to adjust
(highlighted in dark gray).
- Press Orange Button (Edit) to enter adjustment mode.
- Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
- Press Orange Button (Set) to confirm each digit;
repeat for other digits as needed.
- After all digits are set, press Blue Button (Save).
Confirm by pressing Orange Button (Save)
(or Blue Button (NOT) to cancel).
- Press Blue Button (ESC) to return to the main screen (green backlight).
A06 Pipe Thickness
A06 Pipe Thickness Setting Steps
- From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
- Use Gray Arrow Buttons to select A.Measurement, press Orange Button (Enter).
- Use Gray Arrow Buttons to select A06 Pipe Thickness,
press Orange Button (Enter).
- Use Gray Arrow Buttons to choose the digit to adjust
(highlighted in dark gray).
- Press Orange Button (Edit) to enter adjustment mode.
- Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
- Press Orange Button (Set) to confirm each digit;
repeat for other digits as needed.
- After all digits are set, press Blue Button (Save).
Confirm by pressing Orange Button (Save)
(or Blue Button (NOT) to cancel).
- Press Blue Button (ESC) to return to the main screen (green backlight).
A08 Pipe Material
A08 Pipe Material Setting Steps
- From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
- Use Gray Arrow Buttons to select A.Measurement, press Orange Button (Enter).
- Use Gray Arrow Buttons to select A08 Pipe Material,
press Orange Button (Enter).
- Use Gray Arrow Buttons to choose the pipe material.
If no matching option is listed, select custom.
- Press Orange Button (Set) to confirm.
- Press Blue Button (ESC) to return to the main screen (green backlight).
A08 Available Pipe Materials:
3
Probe Installation Method Setting
A27 Wedge Method
A27 Wedge Method Setting Steps
- From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
- Use Gray Arrow Buttons to select A.Measurement, press Orange Button (Enter).
- Use Gray Arrow Buttons to select A27 Wedge Method,
press Orange Button (Enter).
- Use Gray Arrow Buttons to choose the wedge method.
- Press Orange Button (Set) to confirm.
- Press Blue Button (ESC) to return to the main screen (green backlight).
4
Check Probe Spacing for Installation
A29 Probe Spacing Setting Steps
- From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
- Use Gray Arrow Buttons to select A.Measurement, press Orange Button (Enter).
- Use Gray Arrow Buttons to select A29 Probe Spacing,
press Orange Button (Enter).
- Check the required distance between probes shown in the screen
- Record the value to be used during probe installation
- Press Blue Button (ESC) to return to the main screen (green backlight).
5
Fluid Sound Velocity Setting
Z03 Auto Initialize Setting Steps
- From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
- Use Gray Arrow Buttons to select Z.Easy Setup, press Orange Button (Enter).
- Use Gray Arrow Buttons to select Z03 Auto Initialize,
press Orange Button (Enter).
- Press the right gray Start button to begin automatic initialization.
- After completion, press Blue Button (ESC) to return to the main screen
(green backlight).

Troubleshooting
1. Rapid Fluctuating Readings
2. No Zero at Zero Flow
3. Incorrect Positive/Negative Sign Display
4. Unit Adjustment
Last updated