# 4.3 Meter (Measurement Settings)

|                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| ----------------- | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ |
| Damping Factor    | The purpose of setting the damping coefficient is to smooth the output value. 0​ means no smoothing effect. 99​ indicates the maximum smoothing effect. It is recommended to set this factor to 0​ when calibrating the flowmeter.                                                                                                                                                                                                                                                                                                                                                               |
| Low Flow Cutoff   | When the flow rate is set to value output 0​, avoid zero run-in value caused by the invalid cumulative value.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Set Zero          | Due to unavoidable errors in the manufacture of circuits, probes, wires, etc., even when measuring stationary fluids, there is still a time difference. This function can be utilized when the fluid is at a standstill. The system will automatically sample enough zero data to calculate and automatically deduct.                                                                                                                                                                                                                                                                            |
| Reset Zero        | Zero setting value is 0​.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Manual Zero Point | Manually enter the static zero offset value in nanoseconds (ns). The relative flow rate value (in m/s) will be displayed below the input for reference.                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Scale Factor      | Enter this slope correction factor to correct the measured value. For example, if the measured value is 1, but the actual value should be 1.2, then enter a correction factor of 1.2​.                                                                                                                                                                                                                                                                                                                                                                                                           |
| System Lock       | Set a six-digit passcode to lock the settings interface. The password must be input to enter the interface. Enter 0​ to cancel the passcode lock.                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Poor Signal Value | <p>When the ultrasonic signal is too weak or of poor quality, the system will automatically respond in one of the following two ways (default value is 1​): </p><p>1. Hold: Continue using the last measurement data. </p><p>2. Zero: Output is 0</p>                                                                                                                                                                                                                                                                                                                                            |
| Analog Output     | <p>Select 4-20 mA current output items: </p><p>1. Flow Rate </p><p>2. Flow Velocity </p><p>3. Up Temperature </p><p>4. Down Temperature </p><p>5. Sound Velocity </p><p>6. Delta Temperature</p>                                                                                                                                                                                                                                                                                                                                                                                                 |
| 4mA Value         | The corresponding value of the above output item is 4 mA. For example: 4 mA = 0 L/min                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| 20mA Value        | The corresponding value of the above output item is 20 mA. For example: 20 mA = 10 L/min                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| 4mA Cal.Set       | After entering this setting, the analog output terminal will output a 4-mA signal, which can be fine-tuned by setting this parameter. The unit is about 0.004 mA and the adjustable range is -250\~250.                                                                                                                                                                                                                                                                                                                                                                                          |
| 20mA Cal.Set      | After entering this setting, the analog output terminal will output a 20-mA signal, which can be fine-tuned by setting this parameter. The unit is about 0.004 mA and the adjustable range is -250\~250.                                                                                                                                                                                                                                                                                                                                                                                         |
| Rx Peak Level     | This parameter is used as a reference for judging the ultrasonic echo amplitude. The default value is 0.375 V. For example, if the received signal amplitude is 0.3 V, the RSSI will be expressed as 0.3/0.4 \* 100% = 80%. In practice, many factors, such as probe characteristics, wire length, setup, fluid type, etc., will affect peak echoes. Therefore, if setup is good and the Q value of the signal is stable, but the RSSI is too low or too high, this parameter can be modified. Under normal conditions, the following values are recommended: 0.5,​ 0.375​, 0.25​, 0.125​ (V0P). |
| Flow Direction    | <p>Select the flow direction of the current fluid in the pipe (default value is 1​): </p><p>1. Bidirectional: Simultaneous measurement of bidirectional flow. Values will be positive or negative depending on the flow direction. </p><p>2. Forward: Measures forward flow only. Negative flow will show as 0.​ </p><p>3. Backward: Measures negative flow only. Forward flow will show as 0​.</p>                                                                                                                                                                                              |
| Anti-noise        | The default value is 1 ​(ON​). If correct measurements cannot be made due to high flow velocity, set this option to 2 (OFF​) to avoid filtering, and the program will filter out the high flow rate values. Turning off this feature may cause a severe jump in the measured value in rare cases where the flow field of the pipeline is disturbed and the Q value is unstable.                                                                                                                                                                                                                  |