# Z05 Scale Factor

### Definition

This function calibrates the flow meter’s accuracy. If the displayed flow differs systematically from a reference value, the user can set a scale factor to automatically correct the output.

**Usage:**

* The measured flow is multiplied by a fixed factor KKK.
* Range: **+0.001 to +1000**, default **+1.0**.
* Execution automatically updates **K03 (Scale Factor)**.

**Example:**

Final Flow=Measured Flow×K\text{Final Flow} = \text{Measured Flow} \times KFinal Flow=Measured Flow×K

If actual flow = 101.5 L but meter shows 100 L, set K=1.015K = 1.015K=1.015.

{% hint style="info" %}
Z-series parameters currently **do not support Modbus read/write**.
{% endhint %}

### Notes

1. Z05 cannot accept negative values.
2. To measure reverse flow, enable **A17 (Reverse Flow)**.
3. **K03** serves the same function as Z05; users may choose either.
4. (Z-series parameters are for quick setup, consolidating multiple essential settings to simplify installation.)

### Operation Procedure

<details>

<summary><strong>Button Functions (Click to Expand)</strong></summary>

The main unit has **four buttons**. The **bottom line (Line 3)** of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:

* <img src="https://content.gitbook.com/content/i4ECt41Kyy9211OWazj4/blobs/BZuNjMgY75LH0onY5NtL/image" alt="" data-size="line"> **Blue Button -** Leftmost button: Return, exit, or save settings.

Gray Arrow Buttons:

* <img src="https://content.gitbook.com/content/i4ECt41Kyy9211OWazj4/blobs/Br88VpmnPLAPdutxVdZb/image" alt="" data-size="line"> **Left Gray Button:** Page up, increase value, or select item
* <img src="https://content.gitbook.com/content/i4ECt41Kyy9211OWazj4/blobs/3Dzp8nuxmHsaZ7LqBscw/image" alt="" data-size="line"> **Right Gray Button:** Page down, decrease value, or select item
* <img src="https://content.gitbook.com/content/i4ECt41Kyy9211OWazj4/blobs/sq6fX6r39JE0Au2hahAR/image" alt="" data-size="line"> **Orange Button:** Rightmost button: Enter settings, edit, or confirm selection

{% hint style="danger" %}
**Do not hold buttons for extended periods to avoid damage.**
{% endhint %}

{% hint style="info" %}
Button functions may change depending on the operation context to improve usability or prevent accidental presses.
{% endhint %}

</details>

```
Z05 Scale Factor Setting Steps
• From the main screen(green backlight), 
  press Orange Button(Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select Z. Easy Setup , press Orange Button(Enter).
• Use Gray Arrow Buttons to select Z05 Scale Factor, press Orange Button(Enter).
• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).
• Press Orange Button(Edit) to enter adjustment mode.
• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
• Press Orange Button(Set) to confirm each digit; repeat for other digits as needed.
• After all digits are set, press Blue Button(Save).
• Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).
• Press Blue Button(ESC) to return to the main screen(green backlight).
```

<figure><img src="https://3404778090-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2Fi4ECt41Kyy9211OWazj4%2Fuploads%2FR5sffkDG1tnMEbBuQcTc%2FFU-LT-Z05-en.png?alt=media&#x26;token=9775f394-28d3-4ae3-a53e-5d77bc7ecbac" alt="" width="375"><figcaption></figcaption></figure>