Z05 Scale Factor

Definition

This function calibrates the flow meter’s accuracy. If the displayed flow differs systematically from a reference value, the user can set a scale factor to automatically correct the output.

Usage:

  • The measured flow is multiplied by a fixed factor KKK.

  • Range: +0.001 to +1000, default +1.0.

  • Execution automatically updates K03 (Scale Factor).

Example:

Final Flow=Measured Flow×K\text{Final Flow} = \text{Measured Flow} \times KFinal Flow=Measured Flow×K

If actual flow = 101.5 L but meter shows 100 L, set K=1.015K = 1.015K=1.015.

circle-info

Z-series parameters currently do not support Modbus read/write.

Notes

  1. Z05 cannot accept negative values.

  2. To measure reverse flow, enable A17 (Reverse Flow).

  3. K03 serves the same function as Z05; users may choose either.

  4. (Z-series parameters are for quick setup, consolidating multiple essential settings to simplify installation.)

Operation Procedure

chevron-rightButton Functions (Click to Expand)hashtag

The main unit has four buttons. The bottom line (Line 3) of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:

  • Blue Button - Leftmost button: Return, exit, or save settings.

Gray Arrow Buttons:

  • Left Gray Button: Page up, increase value, or select item

  • Right Gray Button: Page down, decrease value, or select item

  • Orange Button: Rightmost button: Enter settings, edit, or confirm selection

triangle-exclamation
circle-info

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

Last updated