D16 Dosing Alarm
Definition
Set the flow meter’s alarm mode when the totalizer reaches the preset target (C04). Default: Do Not Disturb.
Dosing Function (Click to Expand)
1. Flow Meter Dosing Function Description
The dosing function controls a valve based on the flow meter’s totalizer. When the accumulated volume reaches the preset target, the valve closes automatically. After the set buffer time, the totalizer resets to zero, preparing for the next dosing cycle.
This function enables dosing control directly through the flow meter without an external PLC, making it suitable for chemical dosing, liquid replenishment, and batching applications.
2. Parameter Configuration Table
E02
OCT Definition
Sets the control output mode
Must be set to Dosing mode
C03
Dosing Mode
Enables the dosing function
Set to On
C04
Dosing Target
Sets the target volume per dosing cycle
Enter as required (unit: L)
C05
Dosing Reset Time
Sets the buffer time after the target is reached before reset
Set based on system response time (unit: s)
H09
Startup Screen
Sets the main display screen
Set to Dosing mode screen
D16
Dosing Alarm
Alarm indication when dosing is completed
Enable as required or set to Do Not Disturb
3. Dosing Operation
After the valve opens, fluid flow begins and the flow meter accumulates volume in real time. When the accumulated volume reaches the dosing target set in C04, the system closes the valve according to the OCT output logic defined in E02, completing the dosing cycle. A “Dosing Complete” alarm is then triggered based on the D16 setting.
The system then enters the reset delay defined in C05. Once the delay time elapses, the totalizer is automatically reset to zero and the system is ready for the next dosing cycle.

Due to response time delays in valve actuation and control signals, the actual dosing result may differ slightly from the set value. This deviation becomes more noticeable with smaller dosing volumes, higher flow rates, or longer distances between the valve, flow meter, and container.
For higher dosing accuracy, select an appropriate pipe size and piping layout. It is recommended to perform on-site testing and adjust the calibration (scaling) factor as needed before actual operation.
Notes
DND (Do Not Disturb) Mode
System detects abnormalities, but red backlight remains off.
Example: Alarm-A “4 Off Pipe” triggers no backlight.
Abnormalities still appear on the Abnormalities page.
RS485 connection still sends alarm signals.
OCT output continues to signal alarms if configured.
Operation Procedure
Button Functions (Click to Expand)
The main unit has four buttons. The bottom line (Line 3) of the LCD displays the current function of each button. Under normal operation, the button functions are as follows:
Blue Button - Leftmost button: Return, exit, or save settings.
Gray Arrow Buttons:
Left Gray Button: Page up, increase value, or select item
Right Gray Button: Page down, decrease value, or select item
Orange Button: Rightmost button: Enter settings, edit, or confirm selection
Do not hold buttons for extended periods to avoid damage.
Button functions may change depending on the operation context to improve usability or prevent accidental presses.

D16 Options, RS485 Values, Read/Write Commands, and Formats
Off
-
0
On
-
1
Do Not Disturb (Default)
2
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Int16 data format.
Last updated