A09 Pipe Soundspeed
Definition
This function is used to manually set the sound velocity of the pipe, with units set in B04.
Notes
The A09 parameter will only appear if A08 is set to Custom.
Operation Procedure
A08 Pipe Material Setting Steps
• From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select A. Measurement, press Orange Button (Enter).
• Use Gray Arrow Buttons to select A08 Pipe Material , press Orange Button (Enter).
• Use Gray Arrow Buttons to choose the options.
• Press Orange Button (Set) to confirm.
• Press Blue Button (ESC) to return to the main screen (green backlight).
A09 Options, RS485 Values, Read/Write Commands, and Formats
Input Range of This Parameter
Corresponding Values
RS-485 Value
Default
2300
-
Minimum
500
-
Maximum
10000
-
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float 32 data format.
Last updated