H09 Home Page

Definition

Sets the main screen displayed on startup. The FU-LT flow meter offers 9 main screens, with the default being Flowrate (instantaneous flow).

To view or select the startup screen, navigate to HMI → Main Screen.

Notes

N/A

Operation Procedure

Button Functions (Click to Expand)

When the main screen is displayed in green backlight, the four buttons function as follows:

  • Blue Button (leftmost): Return, exit, or save settings.

  • Gray Direction Buttons:

    • Left Gray: Previous Page, decrease value, or select item

    • Right Gray: Next Page, increase value, or select item

  • Orange Button (rightmost): Rightmost button: Enter settings, edit, or confirm selection

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

H09 Home Page Setting Steps
• From the main screen (green backlight), 
  press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select H. System , press Orange Button (Enter).
• Use Gray Arrow Buttons to select H09 Home Page, press Orange Button (Enter).
• Use Gray Arrow Buttons to choose the options. 
• Press Orange Button (Set) to confirm.
• Press Blue Button (ESC) to return to the main screen (green backlight).

H09 Options, RS485 Values, Read/Write Commands, and Formats

Options
Description
RS-485 Numbers

Flowrate (Default)

-

0

Velocity

-

1

Total

-

2

Flowrate (L)

-

3

Velocity (L)

-

4

Total (L)

-

5

Dosing

-

6

Status Page

-

7

Hybrid1

-

8

Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Int16 data format.