H02 LCD Bri.

Definition

Sets the LCD brightness of the flow meter. Three levels are available: High, Medium, Low. Default: Medium.

Notes

N/A

Operation Procedure

Button Functions (Click to Expand)

When the main screen is displayed in green backlight, the four buttons function as follows:

  • Blue Button (leftmost): Return, exit, or save settings.

  • Gray Direction Buttons:

    • Left Gray: Previous Page, decrease value, or select item

    • Right Gray: Next Page, increase value, or select item

  • Orange Button (rightmost): Rightmost button: Enter settings, edit, or confirm selection

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

H02 LCD Bri. Setting Steps
• From the main screen (green backlight), 
  press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select H. System  , press Orange Button (Enter).
• Use Gray Arrow Buttons to select H02 LCD Bri. , press Orange Button (Enter).
• Use Gray Arrow Buttons to choose the options. 
• Press Orange Button (Set) to confirm.
• Press Blue Button (ESC) to return to the main screen (green backlight).

H02 Options, RS485 Values, Read/Write Commands, and Formats

Option
Description
RS-485 Value

Low

-

0

Medium (Default)

-

1

High

-

2

Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Int16 data format.