E12 20mA Fine Adj. uA
Definition
Fine-tunes the flow meter’s 20 mA output value. Unit: µA (microamperes). Default: -200.
Notes
After adjustment and saving, the output current changes simultaneously for real-time verification.
Positive values increase the output
Negative values decrease the output
Operation Procedure
E12 20mA Adj. Setting Steps
• From the main screen(green backlight),
press Orange Button(Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select E. Commute, press Orange Button(Enter).
• Use Gray Arrow Buttons to select E12 20mA Adj., press Orange Button(Enter).
• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).
• Press Orange Button(Edit) to enter adjustment mode.
• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
• Press Orange Button(Set) to confirm each digit; repeat for other digits as needed.
• After all digits are set, press Blue Button(Save).
• Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).
• Press Blue Button(ESC) to return to the main screen(green backlight).
E12 Options, RS485 Values, Read/Write Commands, and Formats
Parameter Input Range
Corresponding Values
RS-485 Value
Default
0
-
Minimum
-2000
-
Maximum
2000
-
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float 32 data format.