D14 Pulse Overlap

Definition

Set the flow meter’s alarm mode for overlapping output signals caused by an excessively long E03 Pulse Width. The alarm code belongs to Alarm – B.

Notes

DND Mode

When Do Not Disturb (DND) mode is enabled:

  • The flow meter detects abnormal conditions but the red backlight will not turn on. Example: Alarm-A “4 Off Pipe” will not trigger the backlight in DND mode.

  • Abnormal conditions are still displayed on the Status/Alarm page.

  • If connected via RS485, the monitoring system still receives the alarm signal.

  • OCT alarm outputs continue to operate if configured.

Operation Procedure

Button Functions (Click to Expand)

When the main screen is displayed in green backlight, the four buttons function as follows:

  • Blue Button (leftmost): Return, exit, or save settings.

  • Gray Direction Buttons:

    • Left Gray: Previous Page, decrease value, or select item

    • Right Gray: Next Page, increase value, or select item

  • Orange Button (rightmost): Rightmost button: Enter settings, edit, or confirm selection

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

D14 Options, RS485 Values, Read/Write Commands, and Formats

Option
Description
RS-485 Value

On

-

1

Do Not Disturb (Default)

-

2

Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Int16 data format.

Last updated