# D02 Pipe Sensor

### Definition

Set the flow meter’s alarm mode when module detect pipe diameter is abnormal. Default: Do Not Disturb. The alarm code belongs to Alarm – A.

### Notes

In **Do Not Disturb (DND) mode**, the screen color does not change during an abnormal condition, but values and status bits on the communication and status pages still indicate the issue.

{% hint style="info" %}

### DND Mode

When **Do Not Disturb (DND) mode** is enabled:

* The flow meter detects abnormal conditions but the red backlight will not turn on.\
  \&#xNAN;*Example:* Alarm-A “4 Off Pipe” will not trigger the backlight in DND mode.
* Abnormal conditions are still displayed on the **Status/Alarm page**.
* If connected via RS485, the monitoring system still receives the alarm signal.
* OCT alarm outputs continue to operate if configured.
  {% endhint %}

<figure><img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2Fyhz46UZUbwHHNzvAmbBg%2FFU-ES-alarm-diagram-a-en.png?alt=media&#x26;token=be074f35-a34e-42da-b381-07dd94a00878" alt=""><figcaption></figcaption></figure>

### Operation Procedure

<details>

<summary><strong>Button Functions</strong> <em>(Click to Expand)</em></summary>

When the main screen is displayed in **green backlight**, the four buttons function as follows:

* <img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2FRnX1fMK1ja5FUx043ZYc%2Fimage?alt=media&#x26;token=f31dd577-ba27-4369-af8b-25ce9afcd1f9" alt="" data-size="line"> **Blue Button (leftmost):** Return, exit, or save settings.
* Gray Direction Buttons:
  * <img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2FR48KXgI2flteQ78PYH8a%2Fimage?alt=media&#x26;token=b5b4ab82-6f88-44c6-8b85-181c7bde6fe9" alt="" data-size="line"> **Left Gray:** Previous Page, decrease value, or select item
  * <img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2FVWLxHCjTZgj8Y3CXUn6X%2Fimage?alt=media&#x26;token=35556f59-f963-4b03-8db1-34f2953ddb58" alt="" data-size="line"> **Right Gray:** Next Page, increase value, or select item
* <img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2F0lsHAW4cWIphk6mKtEoK%2Fbutton-orange.png?alt=media&#x26;token=9168c231-821c-4895-9194-0cf61b2fd06a" alt="" data-size="line"> **Orange Button (rightmost):** Rightmost button: Enter settings, edit, or confirm selection

{% hint style="danger" %}
**Do not hold buttons for extended periods to avoid damage.**
{% endhint %}

{% hint style="info" %}
Button functions may change depending on the operation context to improve usability or prevent accidental presses.
{% endhint %}

</details>

```
D02 Pipe Sensor Alarm Setting Steps
• From the main screen (green backlight), 
press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select D. Alarm & Info, press Orange Button (Enter).
• Use Gray Arrow Buttons to select D02 Pipe Sensor, press Orange Button (Enter).
• Use Gray Arrow Buttons to choose the options. 
• Press Orange Button (Set) to confirm.
• Press Blue Button (ESC) to return to the main screen (green backlight).
```

<figure><img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2FpTg2Tg9CVdtq2cymOnvr%2Ffu-es-D02-en.png?alt=media&#x26;token=1e897afb-75cd-4e77-a9f9-a56859714f82" alt="" width="563"><figcaption></figcaption></figure>

### D02 Options, RS485 Values, Read/Write Commands, and Formats

| Options                               | Description | RS-485 Value |
| ------------------------------------- | ----------- | ------------ |
| On                                    | -           | 1            |
| <p>Do Not Disturb</p><p>(Default)</p> | -           | 2            |

> Refer to the[ Read/Write Memory](https://docs.lorric.com/qr/fu-es-echosense-instruction-manual/communication/modbus-rtu-protocol/read-write-memory-ram) List for the Modbus command format. This parameter uses the Int16 data format.