B05 Dec.Place

Definition

Sets the number of decimal places displayed on the LCD and LED (up to 4 places). Values are truncated (rounded down). Default: 01 (1 decimal place).

Notes

This parameter does not affect calculations; it only affects the display. Cumulative flow is still accumulated using the original data.

Operation Procedure

Button Functions (Click to Expand)

When the main screen is displayed in green backlight, the four buttons function as follows:

  • Blue Button (leftmost): Return, exit, or save settings.

  • Gray Direction Buttons:

    • Left Gray: Previous Page, decrease value, or select item

    • Right Gray: Next Page, increase value, or select item

  • Orange Button (rightmost): Rightmost button: Enter settings, edit, or confirm selection

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

B05 Dec. Place Setting Steps
• From the main screen (green backlight), 
  press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select B.Unit , press Orange Button (Enter).
• Use Gray Arrow Buttons to select B05 Dec. Place, press Orange Button (Enter).
• Use the arrow Buttons to set the number of decimal places:
• Left Gray Button: increase value / Right Gray Button: decrease value
• Press Orange Button (Set) to save.
• Press Blue Button (ESC) to return to the main screen (green backlight).

B05 Options, RS485 Values, Read/Write Commands, and Formats

Option
Description
RS-485 Value

00

Display integer only

0

01 (Default)

Display to 1 decimal place

1

02

Display to 2 decimal places

2

03

Display to 3 decimal places

3

04

Display to 4 decimal places

4

Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Int16 data format.