A23 Zero Offset
Definition
This function performs manual zero-point compensation to correct internal system timing offsets. The setting unit is ns (nanoseconds), not a flow unit.
To configure, refer to the value displayed at the bottom-left of the main screen User Debug Page, and enter that value into A23.
User Debug Page
Make sure H10 Debug Page is turned on.
Return to the main screen and use the arrow keys to navigate to the User Debug Page.
Observe the value in the bottom-left (marked in red). The value may fluctuate slightly; it is recommended to observe for a while and enter the average value into A23.

Notes
N/A
Operation Procedure
A23 Options, RS485 Values, Read/Write Commands, and Formats
Input Range of This Parameter
Corresponding Values
RS-485 Value
Default
0
-
Minimum
-100
-
Maximum
100
-
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float32 data format.