A12 Liquid Soundspeed
Definition
Set the liquid Soundspeed. Default: 1482.
Definition
When the user sets A10 (Fluid Type), A12 (Fluid Sound Speed) will be updated accordingly based on the A10 selection.
The unit of A12 is determined by B04 (Flow and Sound Speed Unit).
Operation Procedure
A12 Liquid Soundspeed Setting Steps
• From the main screen(green backlight),
press Orange Button(Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select A. Measurement, press Orange Button(Enter).
• Use Gray Arrow Buttons to select A12 Liquid Soundspeed, press Orange Button(Enter).
• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).
• Press Orange Button(Edit) to enter adjustment mode.
• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
• Press Orange Button(Set) to confirm each digit; repeat for other digits as needed.
• After all digits are set, press Blue Button(Save).
Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).
• Press Blue Button(ESC) to return to the main screen(green backlight).
A12 Options, RS485 Values, Read/Write Commands, and Formats
Input Range of This Parameter
Corresponding Values
RS-485 Value
Default
1482
-
Minimum
100
-
Maximum
3000
-
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float 32 data format.