A09 Pipe Soundspeed

Definition

This function is used to manually set the sound velocity of the pipe, with units set in B04.

Notes

The A09 parameter will only appear if A08 is set to Custom.

Operation Procedure

Button Functions (Click to Expand)

When the main screen is displayed in green backlight, the four buttons function as follows:

  • Blue Button (leftmost): Return, exit, or save settings.

  • Gray Direction Buttons:

    • Left Gray: Previous Page, decrease value, or select item

    • Right Gray: Next Page, increase value, or select item

  • Orange Button (rightmost): Rightmost button: Enter settings, edit, or confirm selection

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

A09 Pipe Soundspeed Setting Steps
• From the main screen(green backlight), 
  press Orange Button(Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select A. Measurement, press Orange Button(Enter).
• Use Gray Arrow Buttons to select A09 Pipe Soundspeed, press Orange Button(Enter).
• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).
• Press Orange Button(Edit) to enter adjustment mode.
• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
• Press Orange Button(Set) to confirm each digit; repeat for other digits as needed.
• After all digits are set, press Blue Button(Save).
• Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).
• Press Blue Button(ESC) to return to the main screen(green backlight).

A09 Options, RS485 Values, Read/Write Commands, and Formats

Input Range of This Parameter
Corresponding Values
RS-485 Value

Default

2300

-

Minimum

500

-

Maximum

10000

-

Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float 32 data format.