# A06 Pipe Thickness

### Definition

Set the pipe wall thickness. Default values vary depending on the selected pipe unit.

<table data-full-width="true"><thead><tr><th>Pipe OD</th><th>DN15</th><th>DN20</th><th>DN25</th><th>DN32</th></tr></thead><tbody><tr><td><p>Default</p><p>mm</p></td><td>2.7</td><td>2.7</td><td>3.2</td><td>3.2</td></tr><tr><td>inch</td><td>0.106</td><td>0.106</td><td>0.126</td><td>0.126</td></tr><tr><td><p>Minimum</p><p>mm</p></td><td>0.5</td><td>0.8</td><td>1</td><td>1</td></tr><tr><td>inch</td><td>0.02</td><td>0.031</td><td>0.039</td><td>0.039</td></tr><tr><td><p>Maximum</p><p>mm</p></td><td>5</td><td>5.5</td><td>6</td><td>6</td></tr><tr><td>inch</td><td>0.197</td><td>0.217</td><td>0.236</td><td>0.236</td></tr></tbody></table>

<table data-full-width="true"><thead><tr><th>Pipe OD</th><th>DN40</th><th>DN50</th><th>DN65</th><th>DN75</th></tr></thead><tbody><tr><td><p>Default</p><p>mm</p></td><td>3.7</td><td>4.1</td><td>4.1</td><td>5.4</td></tr><tr><td>inch</td><td>0.146</td><td>0.161</td><td>0.161</td><td>0213</td></tr><tr><td><p>Minimum</p><p>mm</p></td><td>1</td><td>1</td><td>1.4</td><td>1.5</td></tr><tr><td>inch</td><td>0.039</td><td>0.039</td><td>0.055</td><td>0.059</td></tr><tr><td><p>Maximum</p><p>mm</p></td><td>6.5</td><td>8</td><td>10</td><td>12</td></tr><tr><td>inch</td><td>0.256</td><td>0.315</td><td>0.394</td><td>0.472</td></tr></tbody></table>

### Notes

1. Pipe size units are determined by **B06 (Pipe Size Unit)**.
2. <mark style="color:red;">Each time the pipe size is changed, please re-execute</mark> <mark style="color:red;"></mark><mark style="color:red;">**Z03 Auto Initialize**</mark> <mark style="color:red;"></mark><mark style="color:red;">and</mark> <mark style="color:red;"></mark><mark style="color:red;">**Z04 Auto Zero Offset**</mark> <mark style="color:red;"></mark><mark style="color:red;">to improve measurement accuracy and ensure stable product performance.</mark>
3. On certain firmware versions, a pipe material confirmation page will appear after setting **A06 Pipe Wall Thickness**. Please select the matching pipe material or choose **Custom**. If a predefined pipe material is selected, **A08 (Pipe Material)** will be synchronized automatically.

### Operation Procedure

<details>

<summary><strong>Button Functions</strong> <em>(Click to Expand)</em></summary>

When the main screen is displayed in **green backlight**, the four buttons function as follows:

* <img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2FRnX1fMK1ja5FUx043ZYc%2Fimage?alt=media&#x26;token=f31dd577-ba27-4369-af8b-25ce9afcd1f9" alt="" data-size="line"> **Blue Button (leftmost):** Return, exit, or save settings.
* Gray Direction Buttons:
  * <img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2FR48KXgI2flteQ78PYH8a%2Fimage?alt=media&#x26;token=b5b4ab82-6f88-44c6-8b85-181c7bde6fe9" alt="" data-size="line"> **Left Gray:** Previous Page, decrease value, or select item
  * <img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2FVWLxHCjTZgj8Y3CXUn6X%2Fimage?alt=media&#x26;token=35556f59-f963-4b03-8db1-34f2953ddb58" alt="" data-size="line"> **Right Gray:** Next Page, increase value, or select item
* <img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2F0lsHAW4cWIphk6mKtEoK%2Fbutton-orange.png?alt=media&#x26;token=9168c231-821c-4895-9194-0cf61b2fd06a" alt="" data-size="line"> **Orange Button (rightmost):** Rightmost button: Enter settings, edit, or confirm selection

{% hint style="danger" %}
**Do not hold buttons for extended periods to avoid damage.**
{% endhint %}

{% hint style="info" %}
Button functions may change depending on the operation context to improve usability or prevent accidental presses.
{% endhint %}

</details>

```
A06 Pipe Thickness Setting Steps
• From the main screen(green backlight), 
  press Orange Button(Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select A. Measurement, press Orange Button(Enter).
• Use Gray Arrow Buttons to select A06 Pipe Thickness, press Orange Button(Enter).
• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).
• Press Orange Button(Edit) to enter adjustment mode.
• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
• Press Orange Button(Set) to confirm each digit; repeat for other digits as needed.
• After all digits are set, press Blue Button(Save).
• Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).
• On certain versions, a “Pipe Material Confirmation” screen will be displayed.
• Use the gray arrow keys to select the matching pipe material or Custom, 
  then press the orange Set key to confirm.
• Press Blue Button(ESC) to return to the main screen(green backlight).
```

<figure><img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2FVyOqKwcHEc7XcknmTyYS%2Ffu-es-A06-en.png?alt=media&#x26;token=92295bd8-96a7-415b-a8dd-b411fe066829" alt="" width="563"><figcaption></figcaption></figure>

## A06 Options, RS485 Values, Read/Write Commands, and Formats

| Input Range of This Parameter | Corresponding Values     | RS-485 Value |
| ----------------------------- | ------------------------ | ------------ |
| Default                       | Refer to the table above | -            |
| Minimum                       |                          | -            |
| Maximum                       |                          | -            |

> Refer to the [Read/Write Memory List](https://docs.lorric.com/qr/fu-es-echosense-instruction-manual/communication/modbus-rtu-protocol/read-write-memory-ram/read-write-memory-read-list) for the Modbus command format. This parameter uses the float 32 data format.