A06 Pipe Thickness
Definition
Set the pipe wall thickness. Default values vary depending on the selected pipe unit.
Default
mm
2.7
2.7
3.2
3.2
inch
0.106
0.106
0.126
0.126
Minimum
mm
0.5
0.8
1
1
inch
0.02
0.031
0.039
0.039
Maximum
mm
5
5.5
6
6
inch
0.197
0.217
0.236
0.236
Default
mm
3.7
4.1
4.1
5.4
inch
0.146
0.161
0.161
0213
Minimum
mm
1
1
1.4
1.5
inch
0.039
0.039
0.055
0.059
Maximum
mm
6.5
8
10
12
inch
0.256
0.315
0.394
0.472
Notes
Pipe size units are determined by B06 (Pipe Size Unit).
Each time the pipe size is changed, please re-execute Z03 Auto Initialize and Z04 Auto Zero Offset to improve measurement accuracy and ensure stable product performance.
On certain firmware versions, a pipe material confirmation page will appear after setting A06 Pipe Wall Thickness. Please select the matching pipe material or choose Custom. If a predefined pipe material is selected, A08 (Pipe Material) will be synchronized automatically.
Operation Procedure

A06 Options, RS485 Values, Read/Write Commands, and Formats
Default
Refer to the table above
-
Minimum
-
Maximum
-
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float 32 data format.