A05 Pipe Perimeter

Definition

Set the pipe circumference. Default values vary depending on the selected pipe unit.

Pipe OD
DN15
DN20
DN25
DN32

Default

mm

69.08

81.64

100.4

119.32

inch

2.720

3.214

3.956

4.698

Minimum

mm

47,1

56.52

75.36

94.2

inch

1.854

2.225

2.967

3.709

Maximum

mm

94.2

109.9

135.02

166.42

inch

3.709

4.327

5.316

6.552

Pipe OD
DN40
DN50
DN65
DN75

Default

mm

150.72

188.4

238.64

279.46

inch

5.934

7.417

9.395

11.002

Minimum

mm

113.04

144.44

182.12

219.8

inch

4.45

5.687

7.17

8.654

Maximum

mm

204.1

244.92

288.88

376.8

inch

8.035

9.643

11.373

14.835

Notes

  1. Pipe size units are determined by B06 (Pipe Size Unit).

  2. Each time the pipe size is changed, please re-execute Z03 Auto Initialize and Z04 Auto Zero Offset to improve measurement accuracy and ensure stable product performance.

  3. On certain firmware versions, a pipe material confirmation page will appear after setting A05 Pipe Wall Thickness. Please select the matching pipe material or choose Custom. If a predefined pipe material is selected, A08 (Pipe Material) will be synchronized automatically.

Operation Procedure

Button Functions (Click to Expand)

When the main screen is displayed in green backlight, the four buttons function as follows:

  • Blue Button (leftmost): Return, exit, or save settings.

  • Gray Direction Buttons:

    • Left Gray: Previous Page, decrease value, or select item

    • Right Gray: Next Page, increase value, or select item

  • Orange Button (rightmost): Rightmost button: Enter settings, edit, or confirm selection

Button functions may change depending on the operation context to improve usability or prevent accidental presses.

A05 Options, RS485 Values, Read/Write Commands, and Formats

Input Range of This Parameter
Corresponding Values
RS-485 Value

Custom

Refer to the table above

-

Minimum

-

Maximum

-

Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float 32 data format.

Last updated