A Measurement Setup

No.

Name

Description

A01

Auto Pipe Spec

Automatically detect pipeline specifications, and automatically jump to A02 to display the results after completion. This function is the same as Z02.

A02

Pipe Type

Select the line model to correct the data. Or select "Custom" to customize the pipe size and material.

(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)

A03

Pipe Geo. Unit

Pipe gauge unit, synchronized with B06.

(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)

A04

Pipe OD

Can be set when A01 is "custom". Dimensions are in accordance with B06. After adjustment, A04 and A06 will be automatically recalculated.

(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)

A05

Pipe Perimeter

Can be set when A01 is "custom". Dimensions are in accordance with B06. After adjustment, A03 and A06 will be automatically recalculated.

(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)

A06

Pipe thickness

Can be set when A01 is "custom". Dimensions are in accordance with B06. A06 will automatically recalculate after adjustment.

(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)

A07

Pipe ID

Can be set when A01 is "custom". Dimensions are in accordance with A02.

(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)

A08

Pipe Material

Set the pipe material, modify the flow data or select "Custom" to customize the sound velocity of the pipe.

(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)

A09

Pipe Soundspeed

When A08 is customized, you can manually set the sound velocity of the pipe, and the unit is set in B04.

A10

Liq. Type

Select the fluid type to correct the data. Or select "Custom" to customize fluid parameters.

A11

Auto Liq. Sound Speed

After enabling this function, the flowmeter will automatically detect the sound velocity of the fluid into the calculation. The default is off. (FW 2.0.0 released.)

A12

Liq. Sound Speed

When A10 is customized, the fluid sound velocity can be manually set, and the unit is set in B04.

A13

Liq. Dyna. Viscosity

When A10 is customized, the fluid dynamic viscosity can be manually set, and the unit is g/cm3.

A15

Damping

Reduce data fluctuations, the lowest is 0, which is off, the default value is 10.

A16

Direction

The direction of the fluid can be reversed, and the normal flow direction can be preset.

A17

Neg Flow (Negative Flow)

Whether to display negative flow, the default is on.

A18

Zero Cutoff

How much flow rate (absolute value) below is displayed as zero, set to 0 means off, default off.

A19

Fast Update

Increased screen update rate to every 0.5 seconds. The default is off.

A20

Anti-Noise

Eliminates unstable noise in measurement, can be turned off, and is turned on by default.

A21

Auto Initialization

Enabling this function will allow the flowmeter to adjust the signal state based on the environmental conditions. Please execute this function once every time you reinstall. This function is the same as Z03.

A22

Auto Zero Offset

Please install the pipeline on a pipeline that is still and full inside, and activate this function to correct the zero flow error. This function is synchronized with Z04.

A23

Zero Offset

Manually adjust or zero the Zero offset data.

A24

Startup Pipe Check

After starting, there will be 5 seconds to confirm whether to use the A01 automatic regulation confirmation function. On by default.