A Measurement Setup
No.
Name
Description
A01
Auto Pipe Spec
Automatically detect pipeline specifications, and automatically jump to A02 to display the results after completion. This function is the same as Z02.
A02
Pipe Type
Select the line model to correct the data. Or select "Custom" to customize the pipe size and material.
(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)
A03
Pipe Geo. Unit
Pipe gauge unit, synchronized with B06.
(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)
A04
Pipe OD
Can be set when A01 is "custom". Dimensions are in accordance with B06. After adjustment, A04 and A06 will be automatically recalculated.
(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)
A05
Pipe Perimeter
Can be set when A01 is "custom". Dimensions are in accordance with B06. After adjustment, A03 and A06 will be automatically recalculated.
(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)
A06
Pipe thickness
Can be set when A01 is "custom". Dimensions are in accordance with B06. A06 will automatically recalculate after adjustment.
(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)
A07
Pipe ID
Can be set when A01 is "custom". Dimensions are in accordance with A02.
(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)
A08
Pipe Material
Set the pipe material, modify the flow data or select "Custom" to customize the sound velocity of the pipe.
(After installation is complete or if the pipeline size or material is changed, please perform Auto Initialize (Z03) and Auto Zero Offset (Z04) again to improve the accuracy of the flow meter readings and ensure product stability.)
A09
Pipe Soundspeed
When A08 is customized, you can manually set the sound velocity of the pipe, and the unit is set in B04.
A10
Liq. Type
Select the fluid type to correct the data. Or select "Custom" to customize fluid parameters.
A11
Auto Liq. Sound Speed
After enabling this function, the flowmeter will automatically detect the sound velocity of the fluid into the calculation. The default is off. (FW 2.0.0 released.)
A12
Liq. Sound Speed
When A10 is customized, the fluid sound velocity can be manually set, and the unit is set in B04.
A13
Liq. Dyna. Viscosity
When A10 is customized, the fluid dynamic viscosity can be manually set, and the unit is g/cm3.
A15
Damping
Reduce data fluctuations, the lowest is 0, which is off, the default value is 10.
A16
Direction
The direction of the fluid can be reversed, and the normal flow direction can be preset.
A17
Neg Flow (Negative Flow)
Whether to display negative flow, the default is on.
A18
Zero Cutoff
How much flow rate (absolute value) below is displayed as zero, set to 0 means off, default off.
A19
Fast Update
Increased screen update rate to every 0.5 seconds. The default is off.
A20
Anti-Noise
Eliminates unstable noise in measurement, can be turned off, and is turned on by default.
A21
Auto Initialization
Enabling this function will allow the flowmeter to adjust the signal state based on the environmental conditions. Please execute this function once every time you reinstall. This function is the same as Z03.
A22
Auto Zero Offset
Please install the pipeline on a pipeline that is still and full inside, and activate this function to correct the zero flow error. This function is synchronized with Z04.
A23
Zero Offset
Manually adjust or zero the Zero offset data.
A24
Startup Pipe Check
After starting, there will be 5 seconds to confirm whether to use the A01 automatic regulation confirmation function. On by default.
Last updated