# Measurement Accuracy

## What is Measurement Accuracy?

Measurement accuracy represents how closely an instrument's reading matches the true value. The higher the accuracy, the closer the flow meter reading is to the actual flow rate.

Accuracy is expressed as a percentage, but there are two different calculation methods — and the practical difference between them is significant.

***

## %FS and %OR — Two Ways to Express Accuracy

### Full-Scale Accuracy (%FS)

%FS calculates the error based on the **maximum measurable value (full scale)**, regardless of the actual flow rate at the time. This means the error value remains fixed and constant across all flow rates.

**Example:** A flow meter with a full scale of 30 L/min and accuracy of ±1% FS:

| Flow                 | Error Value | % of Reading |
| -------------------- | ----------- | ------------ |
| 30 L/min (full flow) | ±0.3 L/min  | ±1%          |
| 3 L/min (low flow)   | ±0.3 L/min  | ±10%         |

At low flow rates, %FS accuracy appears relatively poor compared to the actual reading — this is especially important to note when operating near the lower end of the measurement range.

### Reading Accuracy (%OR / %Rd )

%OR (Of Reading), also referred to as %Rd, calculates the error based on the **actual measured value at the time**, so the error value scales proportionally with the flow rate.

**Example:** A flow meter with accuracy of ±2.5% OR:

| Flow     | Error Value  | % of Reading |
| -------- | ------------ | ------------ |
| 30 L/min | ±0.75 L/min  | ±2.5%        |
| 15 L/min | ±0.375 L/min | ±2.5%        |
| 3 L/min  | ±0.075 L/min | ±2.5%        |

%OR accuracy maintains proportional consistency across the entire measurement range, making it especially meaningful at low flow rates.

***

### Full Scale (%FS) vs Reading Value (%OR) — Illustration

<figure><img src="https://4028902321-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FSBz5Tlpy8GCIkK4ygkgZ%2Fuploads%2F7OjjPn79PhkdpT2TNDN3%2Fimage.png?alt=media&#x26;token=05ff808f-ee74-449a-b28a-528cec44e64f" alt=""><figcaption></figcaption></figure>

***

## FU-ES Measurement Accuracy Specifications

FU-ES uses a **dual-range accuracy specification** — applying a different accuracy expression depending on where the actual flow falls within the measurement range. This design ensures the specification is meaningful across the entire operating range.

| Flow Range               | Accuracy | Method                                                           |
| ------------------------ | -------- | ---------------------------------------------------------------- |
| 10% – 100% of Full Scale | ±3.0% OR | % of Reading — error scales proportionally with actual flow rate |
| 0% – 10% of Full Scale   | ±0.3% FS | % of Full Scale — fixed error value floor                        |

#### **Why use two methods?**&#x20;

At higher flow rates (above 10% of full scale), expressing accuracy as a percentage of reading (±3.0% OR) is both precise and meaningful — the error naturally scales proportionally with flow.

However, at very low flow rates (below 10% of full scale), applying a percentage error to an extremely small value would produce an error value that is too small to reflect actual measurement performance. Using a fixed %FS error floor (±0.3% FS) provides a more realistic accuracy reference at low flow.

#### **Practical implication**

The key is knowing the 10% FS switchover point for your model — this is the boundary between the two accuracy methods.&#x20;

**To find your model's switchover point:**

1. Locate your model in the reference table below.
2. Take the upper end of the recommended flow range as your full-scale reference value.
3. Multiply by 10% — the result is your switchover point.

**Example using FU-ES050 (Full Scale: 1320 L/min):** Switchover point = 10% × 1320 = **132 L/min**

| Flow                | Range            | Applicable Method   | Error Value |
| ------------------- | ---------------- | ------------------- | ----------- |
| 1320 L/min（100% FS） | Above 10% FS     | ±3.0% OR            | ±39.6 L/min |
| 660 L/min（50% FS）   | Above 10% FS     | ±3.0% OR            | ±19.8 L/min |
| 132 L/min（10% FS）   | Switchover point | ±3.0% OR = ±0.3% FS | ±3.96 L/min |
| 66 L/min（5% FS）     | Below 10% FS     | ±0.3% FS            | ±3.96 L/min |
| 13.2 L/min（1% FS）   | Below 10% FS     | ±0.3% FS            | ±3.96 L/min |

> Note that at the switchover point (132 L/min), both methods produce exactly the same error value — this is by design. The dual-range specification is calibrated to ensure there is no jump or gap in accuracy values at the switchover point, making the transition seamless.

***

### Applying to your model

Refer to the table below to find your model's recommended flow range, then calculate the switchover point using the same method.

| Model    | Recommended Flow Range | 10% FS Switchover Point |
| -------- | ---------------------- | ----------------------- |
| FU-ES015 | 1.5 – 120 L/min        | 12 L/min                |
| FU-ES020 | 2 – 200 L/min          | 20 L/min                |
| FU-ES025 | 3 – 300 L/min          | 30 L/min                |
| FU-ES032 | 5 – 470 L/min          | 47 L/min                |
| FU-ES040 | 8 – 770 L/min          | 77 L/min                |
| FU-ES050 | 13 – 1320 L/min        | 132 L/min               |
| FU-ES065 | 22 – 1670 L/min        | 167 L/min               |
| FU-ES075 | 30 – 1870 L/min        | 187 L/min               |

> *Measurement accuracy figures are obtained under controlled laboratory conditions, with fixed pipe and fluid conditions, instruments zeroed prior to testing, and tests conducted at 25°C. Although statistical results were obtained under these conditions, accuracy may vary in customer field environments due to differences in fluid type, pipe condition, installation method, and operating temperature.*
