H03 Green Light
Definition
Sets the duration of the LCD green backlight on the flow meter. Default: Always On.
Notes
In power-saving mode, the LCD still displays the main screen information, but the green backlight is turned off.
Operation Procedure
H03 Green Light Setting Steps
• From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select H. System , press Orange Button (Enter).
• Use Gray Arrow Buttons to select H03 Green Light, press Orange Button (Enter).
• Use Gray Arrow Buttons to choose the options.
• Press Orange Button (Set) to confirm.
• Press Blue Button (ESC) to return to the main screen (green backlight).
H03 Options, RS485 Values, Read/Write Commands, and Formats
Options
Description
RS-485 Value
1 Minute
-
0
Always On (Default)
-
1
10 Seconds
-
2
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Unsigned Int16 (UINT16) data format.