F01 Sim. Volume
Definition
Sets the flow meter’s simulated instantaneous flow. The unit follows the setting in Category B. Default: +60.
Notes
To enable this function, turn on Simulation Mode in F02.
OCT (excluding pulses), RS485, and analog current outputs will operate based on simulated data and totalized flow, affecting both totalized values and historical records.
Operation Procedure
F01 Options, RS485 Values, Read/Write Commands, and Formats
Parameter Input Range
Corresponding Values
RS-485 Value
Default
+60
-
Minimum
-100000
-
Maximum
+100000
-
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the float32 data format.