D03 Bat. Alarm
Definition
Set the flow meter’s alarm mode when CR2032 battery is going to be out of battery. Default: On. The alarm code belongs to Alarm – C.
Notes
Take out the battery and wait one minute, the battery alarm will be displayed.
DND Mode
When Do Not Disturb (DND) mode is enabled:
The flow meter detects abnormal conditions but the red backlight will not turn on. Example: Alarm-A “4 Off Pipe” will not trigger the backlight in DND mode.
Abnormal conditions are still displayed on the Status/Alarm page.
If connected via RS485, the monitoring system still receives the alarm signal.
OCT alarm outputs continue to operate if configured.

Operation Procedure

D03 Options, RS485 Values, Read/Write Commands, and Formats
Off (Default)
-
0
On
-
1
Do Not Disturb
-
2
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Unsigned Int16 (UINT16) data format.