D02 Mag. Alarm
Definition
Set the flow meter’s alarm mode for external magnetic field interference. Default: On. The alarm code belongs to Alarm – A.
Notes
In Do Not Disturb (DND) mode, the screen color does not change during an abnormal condition, but values and status bits on the communication and status pages still indicate the issue.
DND Mode
When Do Not Disturb (DND) mode is enabled:
The flow meter detects abnormal conditions but the red backlight will not turn on. Example: Alarm-A “4 Off Pipe” will not trigger the backlight in DND mode.
Abnormal conditions are still displayed on the Status/Alarm page.
If connected via RS485, the monitoring system still receives the alarm signal.
OCT alarm outputs continue to operate if configured.

Operation Procedure

D02 Options, RS485 Values, Read/Write Commands, and Formats
Off
-
0
On
-
1
Do Not Disturb (Default)
-
2
Refer to the Read/Write Memory List for the Modbus command format. This parameter uses the Unsigned Int16 (UINT16) data format.