Installation and Set-up

  • Verify that the manual version matches the hardware and firmware. To check the manual version, refer to the version page.

  • Ensure the installation environment meets standard measurement requirements. Provide sufficient straight pipe lengths upstream and downstream of the flowmeter to stabilize the flow profile.

  • The optimal installation location is on the side of a horizontal pipe, or on a vertical pipe from downstream to upstream.

  • Confirm the operating temperature, chemical resistance, pipe material and dimensions before installation.

  • Please avoid measuring fluids with magnetic particles or placing the flowmeter near equipments generating strong magnetic field, like motors and electromagnetic valves, which may cause LORRIC’s paddle wheel flowmeter to work improperly, as the paddlewheel uses magnetic flow sensor technology.

Last updated