HMI Overview
Main Unit Functions
The FP-AS510 features Double-screen display:
A 5-digit LED display for high-contrast real-time data readout
A 3-row, 3-color backlit LCD screen for detailed data, page names, units, and button labels
Line 1: Current page name and units
Line 2: Actual measured data
Line 3: Available key functions for the current screen

Button Functions
The main unit has 4 buttons. The bottom line of the LCD (Line 3) shows the function of each button. Typically, the buttons are defined as follows:

LCD Three-Color Backlight Function


The system provides a three-color backlight display. The red backlight is only activated when the alarm setting is “On”; if set to “DND (Do Not Disturb)” or “Off”, the backlight will remain off. The meanings of each color are as follows:
DND (Do Not Disturb) Mode
System detects abnormalities, but red backlight remains off.
Example: Alarm-A “4 Off Pipe” triggers no backlight.
Abnormalities still appear on the Abnormalities page.
RS485 connection still sends alarm signals.
OCT output continues to signal alarms if configured.
Frequency of data update
Due to frequent data updates for real time flow volume, data on the LED screen will update first, then data on the LCD screen. Therefore, they may have about 0.1 seconds updated time difference. between both data sets.If there are not many digits in the data, please use the LED screen for the main data.
Displayed digits on LED screen
Maximum 5 displayed digits are provided on the LED screen. If data digits exceed more than 5 digits, “FFFFF” will be displayed on the LED screen (please refer below image). At this time, for reading complete data, you can try to modify the units or see the 2nd row data on the LCD screen to confirm exact flow volume.

When the flow direction is reversed and the data is negative number, the 1st digit will be a minus sign with four remaining digits.
Last updated