E05 Baud
Definition
Sets the Modbus communication baud rate for the flow meter. The Master and Slave must use the same baud rate for communication. Default: Off.
Notes
This parameter cannot be modified via RS485.
Operation Procedure
E05 Baud Setting Steps
• From the main screen (green backlight),
press Orange Button (Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select E. Commute , press Orange Button (Enter).
• Use Gray Arrow Buttons to select E05 Baud, press Orange Button (Enter).
• Use Gray Arrow Buttons to choose the options.
• Press Orange Button (Set) to confirm.
• Press Blue Button (ESC) to return to the main screen (green backlight).
E05 Options, RS485 Values, Read/Write Commands, and Formats
Options
Description
RS-485 Value
9600
0
19200
1
38400
2
OFF (Default)
4