A04 Damping
Definition
A04 Damping: Smooths flow measurement data. Default: 3.
Higher value → smoother display, smaller fluctuations, slower response.
Lower value → faster response, more sensitive to flow disturbances, larger fluctuations.
Notes
Process Monitoring Application: It is recommended to set A04 higher than the default value to improve display stability and readability, reducing the effect of short-term fluctuations on the displayed value.
Process Control Application: It is recommended to set A04 lower than the default value to minimize measurement delay, allowing flow changes to be reflected in the control system in real time.
Function Description: A04 only affects the smoothness of the display and output; it does not affect the actual measurement accuracy of the flowmeter.
Setting Confirmation: After adjusting A04, it is recommended to observe the system operation for several minutes to verify that the setting meets the application requirements.
Usage Reminder: Setting A04 too high may cause delayed response to flow changes. Adjust according to the actual application scenario.
Operation Procedure
A04 Damping Setting Steps
• From the main screen(green backlight),
press Orange Button(Setup) to enter settings (orange backlight).
• Select A. Measurement by pressing Orange Button(Enter).
• Use Gray Arrow Buttons to select A04 Damping, press Orange Button(Enter).
• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).
• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
• Press Orange Button(Set) to save the adjusted value.
• Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).
• Press Blue Button(ESC) to return to the main screen(green backlight).
A04 Options, RS485 Values, Read/Write Commands, and Formats
Default
3
Minimum
0
Maximum
600