E08 20mA Value
Definition
Sets the value corresponding to 20 mA on the flow meter. Both positive and negative values are allowed. Default: 100.
Notes
The 20 mA value must be greater than the 4 mA value.
Operation Procedure
E08 20mA Value Setting Steps
• From the main screen(green backlight),
press Orange Button(Setup) to enter settings (orange backlight).
• Use Gray Arrow Buttons to select E. Commute , press Orange Button(Enter).
• Use Gray Arrow Buttons to select E08 20mA Value, press Orange Button(Enter).
• Use Gray Arrow Buttons to choose the digit to adjust(highlighted in dark gray).
• Press Orange Button(Edit) to enter adjustment mode.
• Adjust digits: Left Gray Button to increase, Right Gray Button to decrease.
• Press Orange Button(Set) to confirm each digit; repeat for other digits as needed.
• After all digits are set, press Blue Button(Save).
• Confirm by pressing Orange Button(Save) (or Blue Button(NOT) to cancel).
• Press Blue Button(ESC) to return to the main screen(green backlight).
E08 Options, RS485 Values, Read/Write Commands, and Formats
Parameter Input Range
Corresponding Values
RS-485 Value
Default
100
-
Minimum
-100000
-
Maximum
100000
-